9.2
Troubleshooting
SYMPTOM
POSSIBLE CAUSE
REMEDY
The instrument will
not work even though
power is on.
1.
The connection part for
power source may be
loose.
2.
Defective system
electronics
Turn the instrument power off.
Check that the power cord is firmly
plugged into the instrument and
into a live AC electrical outlet then
turn the power ON (AC adapter :
100-240V/ 12VDC 3.33A)
If it is still unable to be operated,
contact your local Customer
Service Office.
Display a message
“Loader error”
The calibration pad is not
fit on the loader.
Replace the strip loader.
System Check values
results show “Fail”
1.
The calibration pad is
not fit on the loader
2.
There is a lot of dirty
contamination
on
the
calibration loader
3.
Unusually calibration
4.
The calibration loader
works incorrectly during
operation
Replace strip loader or Retry to
calibration.
Strip Loading Plate
stops unexpectedly
during operation.
1.Unstable electricity.
2.Measurement
motor
gear or bolster error
Use a rated current.
Contact
your
local
Customer
Service Office if it doesn’t operate
correctly
1.
Can’t
make
program update
2.
Can’t connect to
LIS
3.
When connect to
PC, the items can’t be
shown correctly to
the results
1.
Check the USB Com
Port Number
2.
USB cable between
analyzer
and
PC
is
incorrectly connected
3.
Not recognize USB
Driver
1.
It the set up condition is different
from the left conditions, make it
as same like left side
Check the USB COM Port Number
and then change the USB COM
port Number
2.
Correctly connected. But if it is
not working, replacement
USB Cable(A-B Type).
3.
Install USB Driver