29
Electromagnetic immunity – cont.
The RC-DEX is intended for use in the electromagnetic environment
specified below. The customer or the user of an RC-DEX should as-
sure that it is used in such an environment.
Immunity
Test
IEC 60601
Test level
Compli-
ance
level
Electromagnetic en-
vironment – guid-
ance
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the RC-DEX,
including cables, than
the recommended
separation distance
calculated from the
equation applicable
to the frequency of
the transmitter.
Conducted
RF
IEC
61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
Recommended
separation distance
d = 1.2
√
P
Summary of Contents for RC-DEX
Page 1: ...USER INSTRUCTIONS RC DEX Remote control for Widex tinnitus devices...
Page 6: ...6 Rear view 1 Battery cover 2 Eye for key ring string attachment 1 2...
Page 7: ...7 ACCESSORIES 1 Battery 2 Key ring string 1 2...
Page 39: ...39...
Page 40: ...Printed by HTO 2012 12 9 514 0237 041 01 9 514 0128 041S 03 Manufacturer CIB285 1212...