29
29
450182 • PU[REC] • Technical Reference Manual • Printing version 1.0.0 • November 2
nd
, 2020
MAIN SYSTEM
Strain gauge measurement via
>
Full, half or quarter bridge
>
120 and 350 Ω quarter bridge
>
5 V or 10 V excitation with remote sense
>
Simple connection without soldering
MSI2-STG
Input range
20 mV/V
at 5V excitation
Sensor excitation voltage
5 V or 10 V (±5 V); remote sense support
Maximum current
40 mA per channel
Protection
Continuous short to ground; short circuit limit is 70mA
Supported bridge-types
Full bridge 4 or 6-Wire
Half bridge 3 or 5-Wire
Quarter bridge 3-Wire; 120 Ω and 350 Ω bridge completion
DC accuracy
±0.2 % of reading ±5 μV/V
Bandwidth (-3 dB)
60 kHz
Signal-to-noise ratio; spurious-free SNR;
20 mV/V range
Effective number of Bits; Noise μV/V
PP
SNR
SFDR
ENOB
Noise
PP
Sample rate
[dB]
[dB]
[Bit]
[μV/V
PP
]
5 kS/s
101
124
16.5
0.88
10 kS/s
98
125
16.0
1.4
20 kS/s
95
123
15.5
1.9
50 kS/s
91
120
14.9
3.3
100 kS/s
88
115
14.4
4.5
200 kS/s
86
110
13.9
7
Drift
Offset: 0.4
µ
V/°C; Gain: max 50 ppm°C
Sensor connection
Push-in spring connection; 0.14 to 0.5mm²; AWG 26 to 20
TEDS
For adapter identification and calibration data
The MSI2-STG is designed to connect nearly every strain gauge sensor.
Various bridge-types can be configured by jumper. That makes it very flexible and an ideal solution for strain gauge
measurement on
fixed installations
.
It is also a perfect solution for
harsh electronic environment
. Because this tiny amplifier can be mounted directly next
to the sensor with very short cables in between. The signal is immediately amplified by a factor of 50. This reduces the
impact of electromagnetic disturbances by the same factor. The maximum cable length between MSI and the PUREC is
50 meters.
MSI2-STG
up to 50 m
MSI2-STG
Sensor