![DeVilbiss DV63 Standard Plus Instruction Manual Download Page 61](http://html1.mh-extra.com/html/devilbiss/dv63-standard-plus/dv63-standard-plus_instruction-manual_2489015061.webp)
EN - 61
SE-DV64-HE
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Power Frequency 50/60Hz
Magnetic Field IEC 61000-4-8 30 A/m
30 A/m
Power frequency magnetic fields should be that of a typical location in a typical commercial
or hospital environment.
Conducted RF
IEC 61000-4-6
3 Vrms from
150 kHz to 80 MHz
V1 = 3 Vrms
6 Vrms on ISM & Amateur
Bands
Portable and mobile RF communications equipment should be separated from the device
by no less than the recommended separation distances calculated/listed below:
D=(0.4)
√
P
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
E1 = 10 V/m
D=(0.4)
√
P 80 to 800 MHz
D=(0.7)
√
P 800 MHz to 2.5 GHz
Where P is the maximum power rating in watts and D is the recommended separation
distance in meters.
Field strengths from fixed transmitters, as determined by an electromagnetic
site survey, should be less than the compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment marked with the following
symbol:
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and this device. This
device and system are NOT Life-Supporting
This device is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of this device can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the device as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitter M
150 kHz to 80 MHz
outside ISM bands
D=(0.4)
√
P
80 to 800MHz
D=(0.4)
√
P
800 MHz to 2.5 GHz
D=(0.7)
√
P
0.01
0.04
0.04
0.07
0.1
0.11
0.11
0.22
1
0.35
0.35
0.70
10
1.1
1.1
2.2
100
3.5
3.5
7.0
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Summary of Contents for DV63 Standard Plus
Page 2: ...2 HE SE DV64 HE CPAP 0 45 1 2 5 1 5...
Page 7: ...7 HE SE DV64 HE SmartFlex Bluetooth USB...
Page 8: ...8 HE SE DV64 HE LED 10 2 SD USB...
Page 9: ...9 HE SE DV64 HE 1 5 4 2 SD 6 3...
Page 10: ...10 HE SE DV64 HE 1 2 3 SmartCode Adherence Score...
Page 14: ...14 HE SE DV64 HE 1 2 3 4 100 10 0 2 0 CPAP 1 2 0 0 100 1 2 3 4...
Page 15: ...15 HE SE DV64 HE 1 2 1 24 12 12 24 1 2 3 4 5 0 0 CPAP...
Page 17: ...17 HE SE DV64 HE 1 2 2 BIOS 1...
Page 19: ...19 HE SE DV64 HE DEVILBISS SMARTLINK DeVilbiss SmartLink SmartLink SD SD SD 1 2 1 1 CPAP 2 SD...
Page 21: ...21 HE SE DV64 HE 1 3 2 CPAP 1 CPAP CPAP cmH2O 8 CPAP CPAP lpm 10 psi 50...
Page 23: ...23 HE SE DV64 HE 2 AC AC 1 2 7 1...
Page 24: ...24 HE SE DV64 HE 1 2 1 CPAP 6 10 30 1 2 2 3 10 1 4 5...
Page 25: ...25 HE SE DV64 HE 10 6 3 0 10 30 0 3 6 CPAP 5 CPAP 6 PC WEEE EU 2012 19 X DV6 CPAP...
Page 26: ...26 HE SE DV64 HE 1 1 DC 1 2 2 3 3 1 1 CPAP 1 2 2 3 3 4 Auto START 4 5 5 6 6 1 1 2 2 3 3 4 4...