EN - 50
SE-DV64-HE
TRAVEL INFORMATION
NOTE
– For details on part numbers for alternate power sources, refer to Replacement Items/Spare Parts.
Air Travel
Your device is RTCA/D0-160 Section 21 Category M approved for use on commercial airlines when using battery power. You must turn off the Bluetooth. Refer to Features & Settings -
Add-Ons in the guide for details on Bluetooth.
International Travel
Your device is capable of accepting line voltages of 100-240V~, 50/60Hz. For travel to another country, simply order the correct power cord for that region from your provider or use
proper adapter.
Altitude
This device automatically compensates for altitude between 1060 hPa and 700 hPa (~ 1,400 ft below sea level to 9,800 ft). At higher altitudes, breathing patterns change and you may
lose some benefit of your CPAP therapy.
Battery/DC Operation
Several DC power source options are available:
1. DeVilbiss offers an optional, custom battery (DV6EB) for use as a backup power source when AC power is not available, when traveling by air, or any time portability is desired.
Please refer to the guide included with your battery for details.
2. Your device can be operated from a 12 V DC power source by using the optional DC Cable.
3. Alternately, your device can be powered from a stand-alone battery. Connect to a 12V Deep Cycle Marine Battery using the optional DC power cord and optional DC Battery
Clamp-On Adapter.
4. Lastly, instead of connecting12V DC power directly to the device, you can pass the DC power through an AC Inverter that will deliver AC power to the system. The power rating of
the inverter must be at least 200 watts @ 110V~ / 400 watts @ 220V~.
Approximate Battery Operation Run Time
Example Battery Size = 100 W/Hr
CPAP Only
CPAP with Heated Humidifier
CPAP with Heated Humidifier and PulseDose Module
Heater Setting = 3
Heater Setting = 5
Heater Setting = 3
Heater Setting = 5
CPAP Setting (cmH20)
Minimum Run Time (Hrs)
Minimum Run Time (Hrs)
Minimum Run Time (Hrs)
Minimum Run Time (Hrs)
Minimum Run Time (Hrs)
5
16.9
8.7
5.6
9.5
5.9
10
11.9
7.5
4.7
7.8
5.4
15
8.7
6.4
4.4
7.1
4.7
20
6.8
5.8
4.0
6.1
4.4
NOTE
– Run times are at a breath rate of 20 breaths per minute and standard mask leak.
Summary of Contents for DV63 Standard Plus
Page 2: ...2 HE SE DV64 HE CPAP 0 45 1 2 5 1 5...
Page 7: ...7 HE SE DV64 HE SmartFlex Bluetooth USB...
Page 8: ...8 HE SE DV64 HE LED 10 2 SD USB...
Page 9: ...9 HE SE DV64 HE 1 5 4 2 SD 6 3...
Page 10: ...10 HE SE DV64 HE 1 2 3 SmartCode Adherence Score...
Page 14: ...14 HE SE DV64 HE 1 2 3 4 100 10 0 2 0 CPAP 1 2 0 0 100 1 2 3 4...
Page 15: ...15 HE SE DV64 HE 1 2 1 24 12 12 24 1 2 3 4 5 0 0 CPAP...
Page 17: ...17 HE SE DV64 HE 1 2 2 BIOS 1...
Page 19: ...19 HE SE DV64 HE DEVILBISS SMARTLINK DeVilbiss SmartLink SmartLink SD SD SD 1 2 1 1 CPAP 2 SD...
Page 21: ...21 HE SE DV64 HE 1 3 2 CPAP 1 CPAP CPAP cmH2O 8 CPAP CPAP lpm 10 psi 50...
Page 23: ...23 HE SE DV64 HE 2 AC AC 1 2 7 1...
Page 24: ...24 HE SE DV64 HE 1 2 1 CPAP 6 10 30 1 2 2 3 10 1 4 5...
Page 25: ...25 HE SE DV64 HE 10 6 3 0 10 30 0 3 6 CPAP 5 CPAP 6 PC WEEE EU 2012 19 X DV6 CPAP...
Page 26: ...26 HE SE DV64 HE 1 1 DC 1 2 2 3 3 1 1 CPAP 1 2 2 3 3 4 Auto START 4 5 5 6 6 1 1 2 2 3 3 4 4...