EN - 10
SE-7305
SPECIFICATIONS/CLASSIFICATIONS
Size
H x W x D inches (cm)
7305P & 7305D Series
9.0 x 7.0 x 8.0 (22.9 x 17.8 x 20.3)
Weight
lb. (kg)
7305D Series
3.8 (1.7)
7305P Series
6.3 (2.9)
Electrical Requirements
100-240 V AC 50/60 Hz .75 A max; 12 V DC, 33 W max
Internal Rechargeable Battery
7305P Series
Factory Equipped
7305D Series
Not Factory Equipped (provider-installed option)
Vacuum Range
(
NOTE– Using Flovac Liners may impair the device’s performance.)
7305P & 7305D Series
80 to 550 mm Hg
Air Flow @ pump inlet:
27 LPM (free flow) typical (may be less when running from internal battery)
Container Capacity
7305 D & P Series
800 ml (cc) Disposable
7305 D & P Series - Optional
1200 ml (cc) Reusable
Environmental Conditions
Operating Temperature Range
32°F (0°C) - 104°F (40°C)
Operating Relative Humidity
0-95%
Operating Atmospheric Pressure
9 psi (62 KPa) - 15.4 psi (106 kPa)
Storage & Transport Temperature Range
-40°F (-40°C) - 158°F (70°C)
Storage & Transport Relative Humidity
0-95%
Storage & Transport Atmospheric Pressure
7.3 psi (50 kPa) - 15.4 psi (106 kPa)
Warranty
7305P Series
Two-years limited, excluding internal battery and container
7305D Series
Two-years limited, excluding container
Internal Battery
90-day
Approvals
7305D & 7305P Series
IEC 60601-1:2005+AMD1:2012; EN 60601-1-2: 2015; ISO 10079-1:2015; IEC 60601-1-2:2014; IEC 60529:1989
+AMD1:1999 +AMD2:2013; IEC 60601-1-6:2010; IEC 60601-1-11:2015; IEC 60601-1-12:2015; IP33
CONFORMS TO AAMI STD. ES 60601-1, AAMI STD. HA 60601-1-11, IEC STD. 60601-1-6, IEC 60601-1-12,
ISO STD. 10079-1
CERTIFIED TO CSA STD. C22.2 NO. 60601-1, NO. 60601-1-11, IEC STD. 60601-1-6, IEC STD. 60601-1-12,
ISO STD. 10079-1
7305 Series complies with IEC 60601-1 3rd Edition
Equipment Classifications
With respect to protection from electric shock
Class II and internally powered
Degree of protection against electric shock
Type BF Applied Parts
Degree of protection against ingress of liquids
IP33 and ordinary power supply
Mode of Operation
Intermittent Operation: 30 minutes on, 30 minutes off
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
ISO Classification
7305P Series only - Electrically powered medical suction equipment for field and transport use according to EN ISO 10079-1:2015
High Flow/High Vacuum
7305D Series - Electrically powered medical suction equipment for non-transport use according to EN ISO 10079-1:2015