neoprobe
TM
GDS Operation Manual
AW-000639
Page
7 of 38
Quick Start Probe with cable
Procedure for reusable cabled detector probes (Model 1017 or Model 2059).
Step
Prepare
Prepare console for operation as stated above:
1
If needed, attach external collimator to probe
2
Using aseptic technique (without sterile surgical drape) connect probe to Model 2024 or Model 2060 probe cable;
align red dot on probe to red dot on cable connector; gently insert cable into Probe Input Connector; if using a sterile
surgical drape - place probe in drape now
3
Align arrow on cable connector and groove in Probe Input Connector [5] on console; gently insert cable connector
neoprobe
TM
system is now ready for use
NOTE
: Model 1013 collimator is compatible with Model 1100, 1101, and 1017 probes.
NOTE
: Model 1017 14mm probe connector has four pins; Model 2059 11mm probe connector has six pins.
NOTE
: You hear audible click as you insert cable connector into Probe Input Connector. Connector is push/pull component - DO
NOT TWIST/TURN CABLE CONNECTOR, damage to connectors and connector pins may occur.
3.
DYNAMIC PITCH MODE (DEFAULT)
Radioactive isotopes are traced in Dynamic Pitch Mode from injection site to other areas of concentration.
To….
Perform following…..
And this occurs……
Turn system on
See Equipment Assembly
above
“Dynamic Pitch” indicator [4] is lit
Dynamic Pitch Range automatically set to 0-100
count range, and default background count is 2 cps
Radionuclide automatically set to Technetium 99m
(
99m
TC), and indicator [3] is lit
Select Dynamic Pitch
count range different
from 0-100
Press and release
Dynamic Pitch Count
Range selection buttons
[18].
Indicator for desired Count Range is lit [18];
Background Count Readout [ 8] changes to default
for selected Count Range
Select a radionuclide
different from
99m
TC.
Press and release
Radionuclide Selection
button [21] until desired
radionuclide is lit
Radionuclide Selection Indicator [ 3] is lit
Artwork No: AW-000639 Revision: A Artwork Status: Released Lifecycle Name: Artwork
PCO No: N/A ECN No: ECN-000214 Release Date: 10/09/2012 Effective Date: 10/09/2012