Step 6: 

Rotate the selector knob clockwise to the “HIGH 

FAIL” position.  The green “OPERATOR GROUND” LED 

should turn off and the red LED should illuminate and the 

alarm should sound.

Step 7:

 Disconnect the Monitor Calibration Unit from the 

monitor.

TESTING THE MAT CIRCUIT
Step 1:

 Connect the Calibration Unit’s banana plug wire 

labeled “GROUND” to a known ground.

Step 2:

 

Connect the included banana jack to wire adapter to 

the Calibration Unit’s banana plug labeled “MAT.”  Insert the 

other end of the wire to the monitor’s green terminal block 

labeled “MAT” (see Figure 9).

Step 3:

 Rotate the Calibration Unit’s knob switch to the MAT 

FAIL 10M position.  The monitor’s red “WORKSURFACE 

GROUND” LED should illuminate and the alarm should 

sound.

Note:

 The 

98220

 Monitor Calibration Unit also calibrates the 

following discontinued Desco items:

Full Time Continuous Monitor

19210, 19211, 98210, 98211

Dual Operator Continuous Monitor with Satellites

19208, 19209, 19230, 19231, 98207, 98208

Jewel

®

 Workstation Continuous Mini Monitor

19213, 19214, 19215, 19216, 19217

Multi-Mount Continuous Monitor

19220, 19221, 19222, 19223, 98225, 98226, 98227, 98228

Calibration

Required Test Equipment: RLC Bridge

Settings:

For 50 Hz, Frequency = 1,000 Hz (20 x 50), 20th Harmonic

For 60 Hz, Frequency = 1,020 Hz (17 x 60), 17th Harmonic

Set function switch to read “equivalent parallel circuit”
Additional Required Test Equipment for 

98220

 MAT 

Resistance Measurement:

Megger:  

Set V compliance = 50V or less

or

DMM:   

50V power supply

Record Data for:

Figure 9. Connecting the MAT test lead from the Monitor 
Calibration Unit to the Multi-Mount Continuous Monitor

Step 4: 

Rotate the selector knob counter-clockwise to the 

MAT PASS 10M position.  The monitor’s “WORKSURFACE 

GROUND” green LED should illuminate.

Specifications

Weight:   

6.8 oz

 

 

192 g

Dimensions: 

4.5" x 2.5" x 1.2"

 

 

11.4 cm x 6.4 cm x 3.0 cm

Low Pass

Low Fail

High Pass

High Fail

Serial #

Cp

Dis

Cp

Dis

Cp

Dis

Cp

Dis

Serial #

Mat Pass

Mat Fail

Megohms @ 50V

Compare with the following specs (tolerance = ± 10%):

Equiv. Parallel C

Dissipation Factor

Low Fail

138.9 pF

0.158

Low Pass

118.6 pF

0.367

Hi Pass

49.0 pF

0.445

Hi Fail

44.7 pF

0.192

Pass

Fail

V Measure ~ 50

10 Mat

8 Megohms

12 Megohms

100 Mat 80 Megohms

120 Megohms

500 Mat 400 Megohms 600 Megohms

Mat (tolerance = ± 4%):

Summary of Contents for TB-2082

Page 1: ...brand equipment The following step by step procedures will cover calibration for specific test units The procedures will not cover adjustment of the test equipment For detailed information regarding adjustment of specific Desco brand testers contact our Customer Service Department at 909 627 8178 TECHNICAL BULLETIN TB 2082 Calibrating the Full Time Continuous Monitor The Full Time Continuous Monit...

Page 2: ...uous Monitor Step 3 Rotate the Calibration Unit s knob switch to the MAT FAIL 10M position The monitor s red MAT GROUND LED should illuminate and the alarm should sound Step 4 Rotate the selector knob counter clockwise to the MAT PASS 10M position The monitor s MAT GROUND green LED should illuminate Item Voltage 19232 120 VAC 19233 220 VAC Note If calibrating discontinued items 19208 or 19209 disc...

Page 3: ...nana lead Step 5 Repeat this procedure for the remaining satellite remote Calibrating the Jewel Workstation Continuous Mini Monitor The Jewel Workstation Continuous Mini Monitor is available as following item numbers Figure 5 Connecting the MAT test lead from the Monitor Calibration Unit to the satellite remote Step 3 Rotate the Calibration Unit s knob switch to the MAT FAIL 10M position The monit...

Page 4: ...tion The monitor s red WORKSURFACE GROUND LED should illuminate and the alarm should sound Step 4 Rotate the selector knob counter clockwise to the MAT PASS 500M position The monitor s WORKSURFACE GROUND green LED should illuminate Note During FAIL LOW and FAIL HIGH the monitor will continuously alarm until the banana lead is removed from the satellite remote The alarm will shut off approximately ...

Page 5: ...us Mini Monitor 19213 19214 19215 19216 19217 Multi Mount Continuous Monitor 19220 19221 19222 19223 98225 98226 98227 98228 Calibration Required Test Equipment RLC Bridge Settings For 50 Hz Frequency 1 000 Hz 20 x 50 20th Harmonic For 60 Hz Frequency 1 020 Hz 17 x 60 17th Harmonic Set function switch to read equivalent parallel circuit Additional Required Test Equipment for 98220 MAT Resistance M...

Page 6: ...roximately two weeks If your unit is out of warranty call our Customer Service Department at 909 627 8178 Chino CA or 781 821 8370 Canton MA for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT W...

Reviews: