Desco 19443 Operation & Maintenance Manual Download Page 6

TB-3090

 Page 6 of 8

Positive Discharge Time Measurements

To provide a POSITIVE voltage output, touch the 

plate located underneath the Charger, and press the 

switch forward at the same time. Momentarily touch 

the Charger’s output terminal to the conductive plate 

attached to the Digital Static Field Meter. The Digital 

Static Field Meter reads approxi1.10 kV. By 

using a stop watch or other timing device, determine the 

time needed for the voltages to decrease from +1.10 kV 

to +0.10 kV. This is the positive discharge time.

Negative Discharge Time Measurements

To provide a NEGATIVE voltage output, touch the plate

located underneath the Charger, and press the switch

downward at the same time. Momentarily touch the

Charger’s output terminal to the conductive plate 

attached to the Digital Static Field Meter. The Digital 

Static Field Meter reads 

approximately -1.10 kV. By using a stop watch or other 

timing device, determine the time needed for the 

voltages to decrease from -1.10 kV to -0.10 kV. This is 

the negative discharge time.
NOTE: A ground path must be provided between the 

touch plate of the Charger and the ground reference of 

the Digital Static Field Meter. This is normally provided 

by holding the Charger in one hand and the Digital Static 

Field Meter with Conductive Plate in the other.

Maintenance

The Digital Static Field Meter and Charger are factory 

calibrated and no maintenance is required. Contact 

Desco Customer Service should the products behave 

abnormally. Any unauthorized service will void the 

warranty and result in additional repair charges.
NOTE: This Digital Static Field Meter is a precision 

instrument and should not be subjected to dropping as 

that would void the warranty.

Replacing the Battery

The Digital Static Field Meter’s battery should be 

replaced when a battery symbol appears on the display. 

The battery in the Charger should be replaced annually 

or when it is unable to provide approximately ±1100V. 

Always replace the batteries with a 9V alkaline or 

equivalent battery.

Cleaning

The area around the aperture of the Digital Static Field 

Meter must be kept clean to ensure accurate, drift-free 

readings. Never touch the aperture. Use low-pressure 

instrument-grade air to remove dust or other debris. To 

remove more severe contamination, spray or flush with 

the smallest practical amount of clean technical-grade 

of isopropyl alcohol. Allow the instrument to air dry for 

several hours afterward.
Keep the insulators on the Aluminum Conductive Plate 

clean and free of contaminates that may cause surface 

leakage. To test the performance of the Conductive 

Plate, charge the plate and note the discharge rate in 

a non-ionized area. The self discharge rate to 10% of 

original voltage should be greater than five minutes.

Figure 10. Using the Charger to apply a charge onto the 

Conductive Plate

Test Equipment Depot - 800.517.8431 - 

5 Commonwealth Ave, 

Woburn, 

MA 01801 

 - TestEquipmentDepot.com 

Summary of Contents for 19443

Page 1: ...provides convenience and portability to test per ANSI ESD SP3 3 Periodic Verification of Air Ionizers or Compliance Verification ESD TR53 Desco recommends using the 770004 or 770005 Charged Plate Analyzer if precise measurements are required All Digital Static Field Meters and Air Ionization Test Kits are calibrated to NIST standards and include a certificate When any object becomes electrostatica...

Page 2: ... Static Field Meter features and components C B Figure 3 Air Ionization Test Kit and Carrying Case A Display Displays the measured value in kV inch B Rotary Zero Knob Sets the 0 volt reference point C Power Switch Slide to the left to power the meter off Slide to the right to power the meter on D Measure Hold Pushbutton Switch Press the pushbutton to display live measurements Release the pushbutto...

Page 3: ...Meter s display reads 0 00 Charger Figure 5 Charger features and components A B C D A Output Contact The output contact is connected to an internal power source When the touch plate located underneath the unit is connected to ground the output contact will provide a charge of the indicated polarity The charger is designed so that an operator can press the rocker switch and touch the plate simultan...

Page 4: ...ay and illuminate the red LED located above the pushbutton switch This feature allows the operator to move the Digital Static Field Meter where it may be more easily read or saved for later reference NOTE The LED range indicator will disable while the Digital Static Field Meter is in HOLD mode Use the HOLD feature between measurements to prolong the battery s life Air Ionization Test Kit Performin...

Page 5: ...igital Static Field Meter where it may be more easily read or saved for later reference Performing Discharge Time Measurements In order to verify that an ionizer is operating properly it is also important that its ability to neutralize or discharge static electricity is measured The following procedure will measure an ionizer s discharge time Operating the Charger The Charger has a momentary rocke...

Page 6: ...are factory calibrated and no maintenance is required Contact Desco Customer Service should the products behave abnormally Any unauthorized service will void the warranty and result in additional repair charges NOTE This Digital Static Field Meter is a precision instrument and should not be subjected to dropping as that would void the warranty Replacing the Battery The Digital Static Field Meter s...

Page 7: ...he LCD display reading when depressed General Dimensions 0 9 H x 2 8 W x 4 9 L 24 mm x 70 mm x 126 mm Weight 4 9 oz 140 g with battery Operating Conditions Temperature 50 to 86ºF 10 to 30ºC Relative Humidity To 80 non condensing Altitude To 2000 m Certifications CE Power Requirements Power One 1 9 volt alkaline battery Operating Time Greater than 50 hours with a new battery at 21ºC continuous Air ...

Reviews: