Saftey Instructions

• 

DANGER

 indicates an immediately hazardous 

  situation which, if not avoided, will result in death or 

  serious injury.

• 

WARNING

 indicates a potentially hazardous situation 

  which, if not avoided, could result in death or serious 

  injury.

• 

CAUTION

 indicates a potentially hazardous situation 

  which, if not avoided, will result in property damage.

• 

NOTE

 indicates precautions necessary to avoid 

  potential equipment failure.

This equipment must be installed and maintained as 

outlined in this manual. This equipment is not 

constructed for classified (hazardous) environments. It

cannot be used where it will be exposed to ignitable or 

corrosive materials.

The 

19430

 is intended for use in areas that are free of 

water, oil, solvents and other conductive 

contaminants. Exposure to such contaminants will 

cause failure of the electrical insulation system in the 

product. Formulation of dew on the fieldmeter must be 

avoided.

Do not operate near corrosive fumes of acid/alkali or 

corrosive gases such as chlorine. The 

19430

 must be 

properly grounded to function accurately.

The 

19430

 contains a microcomputer chip. It should not 

be used in an environment with excessive 

electromagnetic noise.  The internal parts of the 

19430

 

should never come in contact with foreign

substances.  

The 

19430

 has been calibrated for a measuring 

distance of 1”± 1/64" (25mm ±0.5mm).

Important Notes and Cautions

1. Store and use the fieldmeter in a clean and dry 

    environment. Foreign substances should never enter 

    or touch the internal parts.

2. Vibration and impact may damage and/or affect the 

    sensitivity of measurement of the instrument.

3. Do not attempt to take apart the fieldmeter.

4. The measuring distance between the fieldmeter and 

    a test object should be 1” ± 1/64” (25mm ± 1mm). 

    The calibration was carried out at this distance. The

    specified accuracy is valid for this distance.

Troubleshooting Guide

Fault

Possible 

Causes

Solution

Fieldmeter is on but

when turned toward a 

charges object, the

number and the bar

graph do not change.

The meter is in HOLD 
mode of operation.

Cancel the HOLD 

mode by pressing the 

HOLD button once.

The display cannot be

set to zero.

The meter is in HOLD
mode.

The true ZERO point 

has shifter.

Cancel the HOLD 

mode by pressing the 

HOLD button once.

If the meter reads 
0.20 or more for an 

uncharged object, a 

read adjustment

may be needed.

Contact your local 

representative or 

Desco Customer

Service for read 

adjustment, repair or 

replacement.

When the POWER 

button is pressed, 

LCD display does not
appear.

No battery, low 

battery capacity or 

bad contact.

Malfunction of the 

LCD display.

Replace the battery / 

Make proper 
connection.

Discontinue use and

contact your local 

representative

or Desco Customer 

Service for read 

adjustment, repair or 

replacement.

A part of the LCD 
display does not 
appear.

Display shows 
ERROR.

Alarm does not 

sound.

Malfunction of the 

LCD display.

Discontinue use and

contact your local 

representative or 

Desco Customer 

Service for

read adjustment, 

repair or replacement.

5. Grounding of the instrument is essential for sensitive 

    ion balance measurements.

6. To maintain the high insulation level of insulation of  

    the plate, the 

19430

 should be stored in a place 

    having less than 60 %RH. It should be kept inside a 

    plastic zipper bag with desiccant.

7. This fieldmeter is not an explosion-proof product. Do 

    not use this product in flammable atmosphere such  

    as paint-thinner, benzene, gasoline, propane, natural 

    gas etc.

8. The instrument should be re-calibrated once a year. 

    Contact Desco Customer Service when a

    recalibration of the fieldmeter is necessary.

Summary of Contents for 19430

Page 1: ...splay Automatic power off after five minutes may be disabled if preferred Figure 1 Digital Static Locator Item 19430 Specifications Measuring range 0 to 1 49 kV Lo range 1 0 kV to 20 0 kV Hi range Range hysteresis 1 0 kV to 1 5 kV 0 to 200 V Ion Balance measuring range Measuring distance 1 1 64 25mm 0 5mm LEDs guide for correct distance between charged object and fieldmeter Response time 1 s LCD d...

Page 2: ... TB 3048 Page 2 of 5 DESCO WEST 3651 Walnut Avenue Chino CA 91710 909 627 8178 Fax 909 627 7449 DESCO EAST One Colgate Way Canton MA 02021 1407 781 821 8370 Fax 781 575 0172 Web Site Desco com circuit This terminal must be grounded properly for accurate measurement 1 Turn the focusing LED and sensor toward an un charged area 2 Press the POWER button A single beep for one second indicates proper op...

Page 3: ...nd to check for residual charges on surfaces The Digital Static Locator can also be used to measure ion balance voltage residual voltage 1 Depress the charge plate release button on either side of the fieldmeter to remove the charge plate assembly 2 Insert the plate assembly into the other end of the fieldmeter the end with the Desco logo until it clicks into place NOTE Ion Balance plate must not ...

Page 4: ...he sensitivity of measurement of the instrument 3 Do not attempt to take apart the fieldmeter 4 The measuring distance between the fieldmeter and a test object should be 1 1 64 25mm 1mm The calibration was carried out at this distance The specified accuracy is valid for this distance Troubleshooting Guide Fault Possible Causes Solution Fieldmeter is on but when turned toward a charges object the n...

Page 5: ...ce or other proof of date of purchase Any unit under warranty should be shipped prepaid to the Desco factory Warranty replacements will take approximately two weeks If your unit is out of warranty call our Customer Service Department at 909 664 9980 Chino CA for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to bring your ...

Reviews: