7 Reprocessing
Operating instructions Rhino XE / K-Style
34
67 16 778 D3686.201.03.02.02
7
Reprocessing
7.1
Automated cleaning
The cleaning equipment used must be approved by its
manufacturer for the cleaning of dental instruments.
For further details, refer to the operating instructions supplied with
the unit.
ü
The instrument is conditioned with a cleaning device.
1.
Check whether the instrument is clean after reprocessing
under good lighting (min. 500 lux) and color rendering index
(min. 80 Ra).
2.
If they are still dirty, repeat the process.
Ä
For further reprocessing, the instrument should be dry and
free of residue.
3.
Blow the instrument out with max. 3 bar.