3 Operation
Page 10
3 Operation
WARNING:
Never plug an instrument on a running dental motor.
The dental motor must be sterilized before first use and after each use.
To connect the Motor to the Motor Housing:
Lightly coat the O-rings on the motor tube with a lubricant spray. Align the pins on the motor with the holes on the motor hose until the
two match, twist the two together to lock as shown below.
Measure the air supply coming out of the end of the motor coupler as shown on the right.
The value must be at least 10 l/min.
WARNING:
Before treatment of the patient, ensure the dental motor is properly connected to the hose by gently pulling on the motor.
To prevent heat buildup on the
patient’s teeth, the water spray must be at least 50 ml/min for appropriate cooling.
To prevent the dental motor from heating up, the air spray must be delivering 10 Nl/min.