SmartLite™ PS
Scientific
Compendium
Fully charged, the battery has a capacity of approx. 150 ten-second cures (approx.
25 minutes operating time). Therefore, under most conditions of use, the device only needs
to be recharged over night.
When the battery light flashes red (fast), the battery needs to be recharged. Another
2 minutes exposure time is then left for completion of the treatment. Light output is not
reduced during this period. After these 2 minutes, the device switches off automatically and
continues to flash red (slowly) indicating the need to recharge.
To maintain the maximum capacity of the rechargeable battery, discharge the device by
regular use at least every 6 months and recharge the battery immediately thereafter.
High voltages are present inside the charger. Do not open charger or the hand piece.
Use only under dry conditions. Do not use if the charger is wet. Store dry. Only authorised
technicians should repair the SmartLite PS LED Tip, hand piece or charger.
11 Guarantee
terms
DENTSPLY grants a 2-year guarantee on hand piece and charger and a 6-month guarantee
on the LED Tip commencing on the date of purchase. Within the guarantee period,
DENTSPLY will eliminate free of charge any defects in the appliance resulting from faults in
material or workmanship either by repairing or exchanging parts or exchanging the whole
device at DENTSPLY's discretion.
21 / 22
Summary of Contents for DeTrey SmartLite PS
Page 1: ......