4 Installation and commissioning
4.4 Initial startup
Dentsply Sirona
Operating Instructions Primescan Connect
30
67 97 216 D3801
D3801.201.01.03.02 2022-09
4.4.1
Plug connections
NOTE
The Primescan scanner is a high-precision optoelectronic scanning
instrument for non-contact impression taking which requires careful
handling. Incorrect handling (impacts, dropping) leads to failure of the
scanner.
➢ Always place the sensitive scanner in its holder!
NOTE
Risk of damage from pulling on the scanner cable
If you pull on the cable itself in order to unplug it or to check the plug
connection, you will damage the cable.
➢ Never pull on the cable.
CAUTION
Placing the scanner holder
The scanner holder must be placed on a flat, horizontal surface inside
or outside of the patient environment. Ensure that the cables of the
scanner do not pose a safety risk.
CAUTION
Trip/fall hazard
The installation of the supply cable (cable between the coupling box
and the scanner) may result in a trip hazard.
➢ Lay the cable so that there is no risk of tripping.
➢ Attach the supply line so that it remains fixed at all times.
WARNING
Hazard for patient and user
If you do not use freely accessible sockets, there is a risk of injury to
patients and users.
➢ Only use sockets which are freely accessible at all times. This
ensures swift disconnection from the power.
CAUTION
Only use the power supply unit supplied by Dentsply Sirona for
operating the coupling box!
CAUTION
Do not connect laptops without approval
Laptops that have not been approved must not be connected. Only
use laptops that are approved according to IEC 62368-1 or IEC
60950-1.