7 Reprocessing
Instructions for Use Midwest Tradition / Tradition Pro / Quiet-Air
46
67 27 155 D3673.201.10.02.02
7.4
Sterilizing
Intervals:
●
Prior to initial operation
●
Prior each subsequent use
Procedure
ü
The high-speed handpiece is cleaned.
ü
High-speed handpiece maintenance is complete.
ü
If necessary, the high-speed handpiece can be sterilized in
packaging suitable for sterilization and storage, e.g. paper/
plastic composite packaging or container.
➢
Sterilize the high-speed handpiece in the steam sterilizer with
saturated water vapor.