Operating Instructions Midwest
®
E Pro / Midwest
®
E Plus / Midwest
®
E mini
7 Conditioning
66 04 206 D3686
.
201.01.04.02
53
Procedure
✔ The spray adapter is disinfected.
✔ The O-ring on the spray adapter is intact.
1. Fit the spray adapter onto the nozzle of the spray can.
2. Insert the instrument until it snaps into place and hold it.
3. Spray the instrument for 1 - 2 seconds.
IMPORTANT: Hold the spray can upright.
4. Wipe any spray that comes out with a disinfection cloth.
5. Repeat the process until the spray is clear.
6604206_EN.book Seite 53 Donnerstag, 15. März 2018 1:08 13