Dentmate LEDEX WL-090 Instructions For Use Manual Download Page 24

Immunity test

IEC60601-1-2
test level

Compliance
level 

Electromagnetic  environment-
guidance

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 Vrms
150 kHz to 
80 MHz
80 %AM(2 Hz)

3 V/m
80 MHz to 
2.5 GHz
80 %AM(2Hz)

3 Vrms

3 V/m

Portable and mobile RF
communications equipment should 
be used no closer to any part of the 
LEDEX

TM 

WL-090, including cables, 

than the recommended separation 
d i s t a n c e   c a l c u l a t e d   f r o m   t h e 
equation applicable to the frequency 
of the transmitter.

Recommend separation distance

d = 1.2 √P 150 kHz to 80 MHz
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz

where P is the maximum output 
power rating of the transmitter 
i n   w a t t s   ( W )   a c c o r d i n g   t o   h e 
transmitter manufacturer and d is the 
recommended separation distance in 
meters (m).

Field strengths from fixed RF

transmitters as determined by an 
electromagnetic site surveyª, should 
be less than the compliance level in 
each frequency range

b

.

Interference may occur in the vicinity 
of equipment marked with the 
following symbol

Note1 : At 80 MHz and 800 MHz, the higher frequency range applies.
Note2 : These guidelines may not apply in all situations. Electromagnetic propagation 

            is affected by absorption and reflection from structures, objects, and people.

1.  Field strengths from fixed transmitters, such as base stations for radio (cellular/

cordless) telephones and land mobile radios, amateur radio, AM and FM radio 
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess 

the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey should be considered. If the measured field strength in the location in which 

the  LEDEX

TM 

WL-090 is used exceeds the applicable RF compliance level above, 

the  LEDEX

TM 

WL-090 should be observed to verify normal operation. If abnormal 

performance is observed, additional measures may be necessary, such as reorienting 
or relocating the LEDEX

TM  

WL-090.

2.  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

21

Summary of Contents for LEDEX WL-090

Page 1: ...Instructions for use...

Page 2: ...to the manufacturing of this product We sincerely hope that it will give you many years of trouble free use Please read and understand all the instructions before using this equipment and save this ma...

Page 3: ...3 Contraindications 11 4 Warnings 12 4 1 User 12 4 2 Ambient conditions 12 4 3 To avoid electric shock shock hazard 13 4 4 Heat development burn hazard 13 4 5 Battery 13 4 6 Accessories 14 4 7 Repair...

Page 4: ...oduct packaging Serial number Catalogue number Manufacturer Date of manufacture Class II AC Adapter Consult the Instructions for use This shows the Type B applied part Do not dispose it with normal ho...

Page 5: ...inferior acrylic that optimizes light conduction and minimizes loss of light from source to tip Therefore it ensures the highest possible intensity of light at the light guide tip Advanced and high e...

Page 6: ...wer supply Input AC100 240 V 50 60 Hz output DC5 V 2 A Optical fiber light guide rod 11 8 mm Filter Anti glare shield Disposable light guide sleeves Disposable curing light sleeves Instructions for us...

Page 7: ...uring mode which is last used a b c a b c a c b d The display allows different information required by the user As shown above it comprises different zones identified from top to bottom as follows A d...

Page 8: ...ich should to be kept during maintenance to prevent the liquid products from damaging the LEDs Then and insert the sterilized light guide into the handpiece Ensure the light guide is properly inserted...

Page 9: ...is charging the display of the handpiece will glow and show the sign of charging battery 2 6 Operating modes Each time disinfect surfaces of the curing light light probes and anti glare cones before...

Page 10: ...2 15 RAMP 5 10 15 20 25 30 35 40 1000 mW cm2 10 STANDARD 5 10 15 20 25 30 35 40 1000 mW cm2 10 HIGH 2 4 6 8 10 1800 mW cm2 10 FAST ORTHO 3 4 5 repeat 10 times 1800 mW cm2 10 TURBO 2 3 4 5 2400 mW cm2...

Page 11: ...e is placed directly over the material in order to be polymerized Extend the curing time accordingly to increase the distance between the light source and the material Recommended curing modes Mode Ap...

Page 12: ...prox half and extend the irradiation time Adaptive Status TURBO 4 A A When LEDEXTM WL 090 is operated frequently for long periods of time the temperature may become too high so the Safety mode functio...

Page 13: ...moisture may cause electrical short circuit or dangerous malfunction 1 The curing time interval should exceed 5 seconds for each cycle 2 Verify the radiometer sensor which can impact the accuracy of...

Page 14: ...th Urticaria solaris or erythropoietic protoporphyria or those who currently have treatments with photosensitising pharmaceuticals 4 While holding the unit steadily press and release the ON OFF Button...

Page 15: ...rby If this happens reduce the interference by reorienting and repositioning the device or screening off the immediate environment The electromagnetic radiation emitted from this device is below the r...

Page 16: ...terrupted Always unplug the charging dock before disinfecting Never use the power supply if the cord has been damaged 4 3 To avoid electric shock shock hazard As it is the case with all high performan...

Page 17: ...es 2401 0005 Using other accessories spare parts may lead to increased emission of electromagnetic interference or to reduced electromagnetic interference immunity 4 6 Accessories 4 7 Repairs and defe...

Page 18: ...show overheat sign If the temperature rises up to a really level please wait a moment and then use it again The intensity is too low If the result of intensity test in Standard mode is under 700 mW cm...

Page 19: ...ly reprocessed and must not be contaminated 9 Warranty DENTMATE TECHNOLOGY Co Ltd warrants the product to be free of manufacturing defects for a period of one year from the date of purchase this is de...

Page 20: ...dimensions 115 max x H68 mm Cradle weight 140 g Equipment class AC Adapter Class II Safety IEC 60601 1 EMC Electro Magnetic Compliance IEC 60601 1 2 Protection from electric shock Type B applied part...

Page 21: ...has been implemented This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices This medical device manufa...

Page 22: ...WL 090 uses RF energy only for internal functions Therefore this RF emission is extremely weak and there is little chance of it creating any kind of interference whatsoever with nearby electronic equi...

Page 23: ...r quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT for 0 5 cycle 5 UT for...

Page 24: ...requency rangeb Interference may occur in the vicinity of equipment marked with the following symbol Note1 At 80 MHz and 800 MHz the higher frequency range applies Note2 These guidelines may not apply...

Page 25: ...W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 kHz to 2 5 GHz d 1 2 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3...

Page 26: ...y 24872 Taiwan TEL 886 2 8976 9226 FAX 886 2 8976 9236 WEBSITE www dentmate com tw EMAIL info dentmate com tw EU representative TOPDENTAL PRODUCTS LTD 12 Ryefield Way Silsden West Yorkshire BD20 0EF U...

Page 27: ...1711 ES V 1...

Reviews: