Electromagnetic Compatibility For Sapphire Plus
WARNING: The Sapphire Plus (and in general all medical electrical equipment) needs to be installed, put into
service, and operated in taking special precautions with respect to EMC (Electromagnetic compatibility).
WARNING: As a preventive measure, separate all electrical equipment from the Sapphire Plus. If space
constraints make this a challenge, determine by observation and testing, if the Sapphire Plus or the other
electrical medical devices are being affected by electromagnetic coupling.
WARNING: the Sapphire Plus is intended for use by Dental professionals only. The Sapphire Plus may cause
Radio Interference. If this happens, take appropriate measures such as re-orienting, relocating or shielding the
Sapphire Plus or filtering the connection to the public mains network.
NOTE: CE marked equipment has been tested and found to comply with the [EN 55011 Class A and EN
60601-1-2:2001].
The limits in these standards are intended and designed to provide reasonable protection against harmful
interference in a typical dental / medical installation. The equipment generates radio frequency energy. If not
installed properly, the Sapphire Plus may interfere with other medical devices. Since there is no guarantee in a
particular installation that other device that interference will not occur, test the installation prior to use. Ensure
that other devices in the vicinity operate without interference and also ensure that the Sapphire Plus is operating
correctly. Test all functions. Correct any interference by one or more of the following mean:
• Reorient or relocate the affected device
• Increase the distance between the devices
• Connect the equipment into an outlet on a different circuit
• Consult the manufacturer or field service technical for help
WARNING: Other mobile or portable RF equipment may impact the Sapphire Plus and other medical
device equipment.
WARNING: Do not modify or use unapproved accessories with the Sapphire Plus. This may increase emissions
or decrease immunity of the Sapphire Plus. If unapproved accessories are used, it becomes the user’s
responsibility to assess EN 60601-1-2 compatibility.
The Den-Mat Holdings, LLC Sapphire Plus Specification
Input Voltage: 100-240 Vac, 50/60 Hz, 6/3A
Wavelength Range: 400-500 nm
Light Intensity (curing mode): 1975 - 2320 mW/cm2 (at time of manufacturing)
Dimensions:131.5 (W) x 234.9 (D) x 263.3 (H) (mm)
Weight: 5.8 kg / 12.8 lbs
Basic System Contents
Sapphire Light Source
Operator’s Manual
Light Guide and Tip
AC Power Cord
Curing Light Protective Glasses (2 pairs)
Composite Test Rings (200 each)
4