Pro-Select Platinum® 

Operation and Instruction Manual

Disclaimers and Special Conditions

•  Equipment is not suitable for use in presence of a flammable anesthetic mixture with air OR with oxygen OR with 

nitrous oxide.

• Equipment is not intended to be defibrillation proof.
•  Some accessories must be sterilized according to common clinical practice. Other accessories may be sterilized 

according to choice. No accessories are supplied in a sterile condition.

•  Equipment is designed for intermittent operation. When using equipment, there is a potential for  

electromagnetic interference with other devices. If there is interference, operate only one of the pieces  
of equipment at a time or physically move them apart until the interference is no longer a problem.

•  There are no waste products or residues that will be generated during the operation of equipment.
•  There are user changeable accessories, but no user serviceable parts in Pro-Select Platinum equipment.
•  There are no special humidity or temperature environmental conditions for transport or storage.
•  Please dispose of any and all component parts of this product according to all local and national  

regulations governing the disposal of electronic devices.

• Service is to be performed by a qualified DenMat professional only.
•  Pacemakers and Ultrasonics - Modern pacemakers are shielded against electromagnetic interference. However, 

most electronic devices do generate an electromagnetic field. Refer to the patient’s cardiologist for approval to 
treat with ultrasonic instrumentation.

Guidance and manufacturer's declaration - electromagnetic emissions

PSP® is intended for use in the electromagnetic environment specified below. The customer or the user of Pro-Select Platinum should assure that it is used in 
such an environment.
                                    Emissions Test                                          Compliance  

Electromagnetic environment - guidance

“RF emissions CISPR 11” 

Group 1

PSP® uses RF energy only for its internal function. Therefore, its RF emissions are 
very low and are not likely to cause any interference in nearby electronic equip-
ment.

“RF emissions CISPR 11”

Class A

PSP® is suitable for use in all establishments other than domestic and those 
directly connected to the public low voltage power supply network that supplies 
buildings used for domestic purposes.

“Harmonic emissions IEC 61000-3-2”

Not applicable

“Voltage fluctuations/  

flicker emissions Complies IEC 61000-3-3”

Complies

Guidance and manufacturer’s declaration electromagnetic emissions 

PSP® is intended for use in the electromagnetic environment specified below. The customer or the user of Pro-Select Platinum® should assure that it is used 
in such an environment.

 

Emissions Test 

IEC 60601 test level

Compliance

Electromagnetic environment - guidance

Electrostatic discharge (ESD) IEC61000-4-2 

±6kVcontact ±8kVair 

A

Floors should be wood, concrete, or ceramic tile. If floors are 
covered with synthetic material, the relative humidity should be 
at least 30%. 

Electrical fast transient/burst IEC61000-4-4 

±2kVforpowersupply lines 

±1kVforinput/output lines 

A

Mains power quality should be that of a typical commercial or 
hospital environment.

Surge IEC61000-4-5

±1kVdifferentialmode 

±2kVcommonmode 

A

Mains power quality should be that of a typical commercial or 
hospital environment.

Voltage dips, short interruptions and voltage 

variations on power supply input lines IEC61000-

4-11

<5%UT(>95%dipinUt)  

for 0.5 cycle 

<40%UT(60%dipinUt ) 

for 5 cycle 

<70%UT(70%dipinUt ) 

for 5 cycle 

<70%UT(70%dipinUt ) 

for 5 cycle 

A, C

Mains power quality should be that of a typical commercial or 
hospital environment. If the user of Pro-Select Platinum requires 
continued operation during power mains interruptions, it is 
recommended that Pro-Select Platinum be powered from an 
uninterruptible power supply or a battery.

Power frequency (50/60Hz)magnetic field 

IEC61000-4-8

3A/m

A, C

Power frequency magnetic fields should beat levels characteristic 
of a typical location in a typical commercial or hospital 
environment.

NOTE:  UT is the a.c. mains voltage prior to application of the test level. Compliance level A: Provides essential performance and remains safe. Compliance level 

C: Loss of essential performance provided the Equipment Under Test remains safe, experiences no component failure and is restorable to the pre-set state 
with operator intervention.

Summary of Contents for PSP Pro-Select Platinum

Page 1: ...1 Pro Select Platinum Operation and Instruction Manual Operation and instruction manual...

Page 2: ...nts Two instruments in one ultrasonic scaling and heated irrigation system PSP is a superb piezo ultrasonic scaling solution and the only system that combines scaling with heated subgingival irrigatio...

Page 3: ...on 10 Operation 11 Foot pedal Scaler mode Irrigation mode Cleaning Disinfection and Sterilization 13 Control unit Bottles Cables Handpieces Tips LED lights Maintenance and Service 15 Control unit Bott...

Page 4: ...Instruction Manual 2 Product Documentation Date installed ___ ___ _______ Installer Control Unit Handpiece Irrigator Tips Tip Date of first use Handpiece cable Cable Date started Date finished LED lig...

Page 5: ...s Power button Handpiece cable connector LCD status display Brackets Handpiece holder Light button Bottle select button Flow rate button Irrigation button Adjustment buttons Serial number Fuse Power t...

Page 6: ...s will void warranty Scaler handpiece S40801 Connect handpiece to LED cable and attach tip see page 8 Irrigator handpiece S40851 Connect handpiece to LED cable and attach 23g cannula see page 8 Foot p...

Page 7: ...e tip wear Tip grip 100540 Non slip rubber makes tip changing easier 2 0 NaF 2037MTM Increased patient comfort because no trays are required In office rinse compatible with composite and porcelain den...

Page 8: ...le materials Do not use this product in presence of inflammable anesthetics or gases Use recommended fuse 1 Amp 5x20 Metric medium time delay Attention During vibration handpiece and cable may affect...

Page 9: ...move depress locking tab and remove connector from jack Plug in power transformer 1 Plug gray transformer connector into gray jack on control unit back 2 Push until it clicks 3 Control unit will beep...

Page 10: ...n the top of scaler or irrigator handpiece ring then clip to the bottom of handpiece Connect handpiece cable 1 Plug handpiece cable into cable connecter ensure black ring is seated in opening 2 Plug S...

Page 11: ...h fingers Connect irrigator tip Warning Irrigation handpiece heater has been designed for use with a 23 gauge side ported cannula Do not use other gauge cannulas as fluid heat variation will result 1...

Page 12: ...se power level Press or press Press or press Adjust scaler fluid flow rate Press then use to increase flow and to decrease flow Select solution bottles Press and hold down outer pedal roll foot to cen...

Page 13: ...the effective working area When scaling use light sweeping eraser like strokes Do not tap tooth or use a start stop motion For best results use 3 5 grams of lateral pressure The round tip profile all...

Page 14: ...ep will sound 5 Place cannula on irrigator handpiece 6 While waiting for heating cycle activation bend cannula as appropriate for access 7 Select desired fluid bottle Use or press on control unit A do...

Page 15: ...ons After each patient medicament use and at end of day each bottle port on control unit handpiece cable and handpieces should be flushed with distilled water for 30 seconds Place distilled water fill...

Page 16: ...with pH neutral surface disinfection solution Attention Do not use disinfection solution containing phenols Do not spray cleaning solution directly on the handpiece 2 Remove LED light from handpiece...

Page 17: ...through the hole then the tip is no longer functional and needs to be replaced 3 Regularly inspect for bent tips or wear patterns that result in flattening of the side s of the tip Tips that exhibit...

Page 18: ...hnical support team can repair the unit and test it for functionality prior to returning to the customer Hardware may only be returned to the company after a Return Material Authorization RMA number h...

Page 19: ...rking Make sure irrigator assembly set up has been precisely followed Make sure system is in irrigator mode Make sure irrigator handpiece is being used Make sure cable is firmly attached to handpiece...

Page 20: ...e tips S30960 Set of three universal straight tips S40200 Solution bottle with lid assembly S40750 LED handpiece cable S40871 LED light S40801 Scaler handpiece S40851 Irrigator handpiece S40901 Multi...

Page 21: ...t RF emissions CISPR 11 Class A PSP is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power supply network that supplies buildings us...

Page 22: ...ributed by DenMat 2014 Den Mat Holdings LLC All rights reserved PSP Pro Select Pro Select Platinum and Pro DenRx are registered trademarks 823060400 09 14BA 1017 W Central Ave Lompoc CA 93436 1 800 43...

Page 23: ...ributed by DenMat 2014 Den Mat Holdings LLC All rights reserved PSP Pro Select Pro Select Platinum and Pro DenRx are registered trademarks 823060400 09 14BA 1017 W Central Ave Lompoc CA 93436 1 800 43...

Page 24: ...Mat PSP Pro Select Pro Select Platinum and Pro DenRx are registered trademarks 2020 Den Mat Holdings LLC All rights reserved 823060400 R 2 02 20CM 800 433 6628 denmat com CONFORMS TO UL STD 60601 1 CE...

Reviews: