Warnings

Battery

•  Inspect batteries prior to use.
•  If the battery is dropped or receives a mechanical impact, ensure that the casing is not damaged.
• Do not use damaged batteries.
•  The electrolyte contained within the lithium cells can cause irritation to the respiratory tract, eyes and skin. In      

addition, cell venting could result in air contaminants, including corrosive or flammable vapors. All precautions 
should be taken to limit exposure to the electrolyte vapor.

Light

•  Turn the light off when not in use. The lamp housing can get hot.
•  Do not use the LED lamp if ambient temperatures exceed 35°C (95°F).
•  When light is in use, always keep the front of the light at least 6 inches from skin and any surface being illuminated.
• Do not shine light into patient’s eyes.
•  Do not shine light onto combustible material. 
• Do not allow contact of the lamp to the patient.
•  Use a curing filter when working with any light-curable resins.

Charging

•  When charging the battery, inspect the charging lights to ensure that charging is active.
•  Do not simultaneously touch the charging unit (when plugged in) and the patient.
•  Use only PeriOptix charging station to charge the batteries. Li-ion batteries can be dangerous and even explode if 

improperly charged. Batteries are designed to be charged only by the supplied charging station.

•  Do not position the battery charger so it is difficult to disconnect the power adapter from the wall.

General

•  Check all components for burrs, nicks, and sharp edges prior to use.
•  If unit is jarred or dropped, re-inspect prior to use.
•  Do not use the PeriOptix system or charge the battery in an oxygen-rich environment.
•  Do not store liquids above the charger, power pack or any other PeriOptix component or accessory.
•  If a spill occurs into the charger, battery, LED lamp or power pack, discontinue use. Using the charger after liquid is 

spilled into the unit could result in electrical shock.

• Clean only as directed.
•  Portable RF communications equipment (including peripherals such as antenna cables and external antennas) 

should be used no closer than 30 cm (12 inches) to any part of the PeriOptix system, including cables specified by 

the manufacturer.  Otherwise, degradation of the performance of this equipment could result. 

•  Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this 

equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this 

equipment and result in improper operation. 

• Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in  

   improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify  

   that they are operating normally.
• Removing the power adapter from the wall outlet is the means of isolating the Solaris

®

/Miniheadlight from the     

   supply means.
• Compliance with radiated emissions and immunity Standards:  Complies with RF Emissions test per CISPR  

   11 Group 1 Class B; and Immunity test per IEC 60601 test level for ESD, EFT, Surge, Power Frequency, Voltage,  

   Conducted RF and Radiated RF.

Warranty

PeriOptix, Inc. warrants its products to be free of original defects in material and workmanship and to perform in 

accordance with specifications for the following items:

Limited Lifetime Warranty

 

All working parts of the Loupe hinge mechanism.

One-Year Limited Warranty  

All components of Solaris and Mini headlight system.

Warranty Exclusions:

•  Breakage or failure due to tampering, misuse, neglect, accidents, modification or shipping.
•  If the instrument is not used in accordance with manufacturer’s recommendations or instructions.
•  If repaired or serviced by other than PeriOptix or a PeriOptix authorized representative
Prompt inspection and reporting of missing or damaged product should be reported to shipping carrier and    

PeriOptix within 3 days of receipt of product.
If PeriOptix

®

 products or any component thereof is found to be defective or at variance with the instrument or com-

ponent(s) it will be replaced at no cost to the purchaser. This warranty only applies to products purchased new from 

PeriOptix or its authorized distributors or representatives.
For any questions or problems, please contact your local distributor or call DenMat Customer Care at 800-433-6628 

between the hours of 8am to 5pm (Pacific Time) Monday-Friday. Outside the U.S., please call 1-805-347-7990 or 

visit denmat.com

©2019 Den-Mat Holdings, LLC. All rights reserved. Surgitel is a registered trademark of General Scientific Corp. Orascoptic is a registered trademark 

of Surgical Acuity, Inc. Zeiss is a registered trademark of Carl Zeiss AG, Corp. CaviCide and CaviWipes are registered trademarks of Metrex 

Research, LLC.  750195000 R:3 02/19CM

Den-Mat Holdings, LLC
1017 W. Central Avenue
Lompoc, CA 93436 USA

denmat.com

MDSS GmbH

Schiffgraben 41

30175 Hannover, Germany

Reviews: