Rev. 02/01/18
VER SMS-YS0320140828-EN
10
low speed)
4. Paste a spot of conducting glue or toothpaste on
the contact interface (1/3 slice side) between test
electrode and selected tooth.
Then press the power switch.
Then lay the test electrode on the on the surface of
selected tooth. Afterwards, the unit will be activated and
simultaneously the figure keeps rising on the screen.
5. When the patient slightly feels toothache, or
anesthesia, you should take the test electrode away from
the tooth and observe the figure on the screen in order to
record it; this figure is the tooth's stimulation current
reaction number.
6. Peak of stimulus current reaction numerical value is
80. Between 0-40, the patient who has the reaction of
ache and anesthesia, the dentist can be sure that the
nerve is still alive. Only the numerical value go up to 40-80,
the patient has the above-mentioned reaction, the dentist
can be sure that part of the teeth nerve has been dead!
When the numerical value has reached 80, but the tooth
has no above-mentioned reaction, this shows the nerve
has already been dead!
7. After operation, the measuring result remains frozen