Rev. 02/01/18
VER SMS-YS0320140828-EN
USER’S MANUAL
DY310
Pulp Tester
The unit must be installed by a qualified dental engineer.The unit is only for use by dental professionals.
Read this operation manual carefully before operation.
Page 1: ...R SMS YS0320140828 EN USER S MANUAL DY310 Pulp Tester The unit must be installed by a qualified dental engineer The unit is only for use by dental professionals Read this operation manual carefully be...
Page 2: ...CAL INDEX SECTION VI OPERATION SECTION VII SAFETYPRECAUTIONS SECTION VIII MAINTENANCE SECTION IX TOUBLESHOOTING GUIDE SECTION X PACKING LIST SECTION XI STORAGE TRANSPORT ENVIRONMENT SECTION XII WARRAN...
Page 3: ...d by the approbatory technician by our company or obtained related nation quality level license professions 2 The power supply shall be in conformity with the relevant provisions of the state and the...
Page 4: ...EMMENDINGEN GERMANY SECTION II DESCRIPTION Pulp Tester DY310 is a device to examine the vitality of dental pulp using the electrical stimulation During the pulp test the current stimulates intradental...
Page 5: ...Rev 02 01 18 VER SMS YS0320140828 EN 4 SECTION III COMPONENTS AND FEATURES A Test electrode B Stainless hook C Control part D Probe cable Test electrode 2pcs Stainless hook 4pcs Probe cable 1pc...
Page 6: ...Rev 02 01 18 VER SMS YS0320140828 EN 5 Press ON OFF button for 2S to turn on the device Press MODE button to set the speed of number change...
Page 7: ...atient comfort If the patient indicates perception simply release the button The stimulus stops immediately but the numbers remain frozen on the face for about 3 minutes Convenient to operate Peak of...
Page 8: ...this manual on the label or on it s accessories Some of the symbols represent standards and compliances associated with the endo motor and its use Consult accompanying documents Cautions Authorized Re...
Page 9: ...uction manual booklet Direct current The device should not be used after the end of the shown or the day DISPOSAL Do not dispose this product as unsorted municipal waste Collection of such waste separ...
Page 10: ...ice 2 Strictly separate the tooth which will be measured from the saliva blow the surface of the tooth until it is dry so as to forbid the stimulation electric current to conduct from the gum or else...
Page 11: ...de away from the tooth and observe the figure on the screen in order to record it this figure is the tooth s stimulation current reaction number 6 Peak of stimulus current reaction numerical value is...
Page 12: ...e charged SECTION VII SAFETYPRECAUTIONS a Before operation you have to read this usage manual carefully b CAUTION Persons having a history of photosensitive reactions or who are using photosensitizing...
Page 13: ...y this product oversea g CAUTION Persons having a history of photosensitive reactions or who are using photosensitizing drugs should not be tested by pulp tester h CAUTION Only experienced professiona...
Page 14: ...unit should be scrubbed by pure water or ethanol and follow the standard disinfection procedure to disinfect the materials Then the pulp tester should be placed in the original packing box in a dry a...
Page 15: ...good contact with cable socket of control part Make sure that all connection cables are in good order Make sure that test electrode connection is well and stable The front part of test electrode can...
Page 16: ...AND SHIPPING CONDITIONS Ambient temperature 40 55 Relative humidity range 80 Atmospheric pressure 50kPa 106kPa Equipment is not suitable for storage in the presence of sunlight rain dust corrosive ga...
Page 17: ...rantee for accessories is 6 months All accessories of the device are damaged or needed to be renewed the user can purchase new accessories from the seller WARNING The device is not repairable by user...
Page 18: ...is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The DY310 uses RF energy only for its internal function Therefore its RF em...
Page 19: ...wing warning is heeded Warning This equipment system is intended for use by healthcare professionals only This equipment system may cause radio interference or may disrupt the operation of nearby equi...
Page 20: ...uld assure that it is used in such an environment Immunity Test IEC 60601 Test level Complian ce level Electromagne tic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV...
Page 21: ...t of a typical commercial or hospital environment The electrical fast transient burst EFT is generated by the switching of inductive loads Separation between the equipment and other loads shall be con...
Page 22: ...UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycle 70 UT 30 dip in UT for 25 cycle 5 UT 95 dip in UT 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycle 70 UT 30 dip in UT for 25...
Page 23: ...magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment NOTE UT is the a c mians voltage prior to application of the test level Table...
Page 24: ...netic environment guidance Condu cted RF IEC 61000 3V rms 150KHz to 80MHz 3V Portable and mobile RF communications equipment should be used no closer to any part of the DY310 including cables than the...
Page 25: ...the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromag...
Page 26: ...io AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey shou...
Page 27: ...help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the DY310 as recommended below according to the ma...
Page 28: ...to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation dista...
Page 29: ...v 02 01 18 VER SMS YS0320140828 EN 28 WARRANTY REGISTRATION FORM Item Name Model Name Serial No Date of Purchase Name Address Phone Email Name of Distributor Authorized Distributors Stamp and Signatur...