3.
To select the
Temperature Scale
(Celsius or Fahrenheit), insert a 3mm
screwdriver (or other tool) into the
Mode Switch Button Hole
on the
back of the device. (Figure 8)
4.
Press the screwdriver (tool) into the hole to select the desired
Temperature Scale
. (Figures 9 & 10)
5.
To select the
Measurement Mode
(Surface or Body) or
Memory
, insert
0.1in (3mm) screwdriver (or other tool) into
Mode Switch Button Hole
,
press and hold for 3 seconds.
6.
Continue pressing the screwdriver (tool) into the hole to select the desired
Measurement Mode
(Sur for Surface; bod for Body) or
Memory
(Cou).
(Figures 11, 12 & 13)
Installing/Replacing Battery
1.
Remove the screws holding the battery compartment cover in place.
2.
Remove the battery compartment cover. (Figure 14)
3.
Insert one 18650 Lithium-Ion battery, observing correct polarity. (Figure 15)
4.
Place cover over battery compartment and re-insert screws.
The device includes a 18650 rechargeable Lithium-ion battery with overcharge
protection. Before installing, please charge the battery for 4 hours to ensure
full power. Due to freight regulations, Lithium-ion rechargeable batteries may
not be fully charged prior to shipping. When the original battery is depleted,
please ensure the replacement also includes overcharge protection.
Fig. 9
Fig. 10
Fig. 8
Fig. 13
Fig. 12
Fig. 11
Fig. 14
Fig. 15