97
│
www.deltaww.com
User Manual
A9000
AC Power Source
[SOURce:]PULSe:FREQuency
Function
:
Set (query) frequency of PULSE waveform.
Example
:
[SOURce:]PULSE:FREQuency?
[SOURce:]PULSE:FREQuency 50.0
Parameter
:
<NR2> Effective range
:
30.0 ~ 1000.0 (unit: Hz)
Parameter returns
:
<NR2>
[SOURce:]PULSe:SPHase
Function
:
Set (query) start phase angle of PULSE waveform.
Example
:
[SOURce:]PULSE:SPHase?
[SOURce:]PULSE:SPHase 60.0
Parameter
:
<NR2> effective range
:
0.0~ 359.9
Parameter returns
:
<NR2>
[SOURce:]PULSe:COUNt
Function
:
Set (query) how many times to execute PULSE voltage.
Example
:
[SOURce:]PULSE:COUNt?
[SOURce:]PULSE:COUNt 500
Parameter
:
<NR1> Effective range
:
0 ~ 10000
Parameter returns
:
<NR1>
[SOURce:]PULSe:DCYCle
Function
:
Set (query) execution time of PULSE waveform, it must be no longer than
total period in PULSE mode.
Example
:
[SOURce:]PULSE:DCYCle?
[SOURce:]PULSE:DCYCle 300
Parameter
:
<NR1>, effective range
:
1 ~ 15999999 (unit: msec)
Parameter returns
:
<NR1>