Delta 62 TP3 Series Installation And Setup Instructions Download Page 16

Page 16

209989   Rev. B

w w w . s p e c s e l e c t . c o m

TROUBLESHOOTING GUIDE

Restart controller by disconnecting the power connector, wait 10 seconds and reattach.  The controller will go through its normal

startup routine.

If the faucet is not working properly, first check for error codes by doing the following:
 

Press and hold TOGGLE (white pushbutton on right hand side) button for 5 seconds.  Once you see the “E”, press TOGGLE

 

button to cycle through and note the error codes:

  ●  

E01/E02 - Lost /Poor Communication with Sensor

                               - Confirm sensor is connected and wire harness is not damaged.

                               - Confirm connectors are clean, dry and corrosion free.

                               - If extension is used, ensure both ends are connected.

                               - Restart controller and retest.

  ●  

E03 - Controller Problem

                                - Restart controller and retest.  If failure continues, contact Technical Service.

  ●  

E04 - Proximity Calibration Failure

                                - Confirm sensor module is connected to faucet shank.

                                - Ensure faucet and sensor module is electrically isolated from sink and any other conductive materials.

                                - Route sensor cables away from power supply wires of any type.

                                - Check grounding (see section on Grounding).

                                - For surface mount box #0, ensure polymer braided hose is used on both the inlet and outlet of the solenoid.

  ●  

E05 - Solenoid

                                - Confirm solenoid is connected.

                                - If possible, cross test solenoid and controller with a known functional controller and solenoid.

  ●  

E06/E07 - Low Power

                                - On battery product, replace batteries, be sure to use the supplied battery holder or repair part.

                                - On hardwired product, check supply voltage (see below).

This Delta Commercial faucet is designed and engineered in accordance with the highest quality and performance standards. With proper

care, it will give years of trouble free service.  Care should be given to the cleaning of this product.  Although the chrome finish is extremely

durable, it can be DAMAGED by ACIDIC CLEANERS (i.e. cleaners designed specifically for vitreous china lavatories and water closets),

HARSH ABRASIVES or POLISH.  To clean, simply wipe the surface with a damp cloth and blot dry with a soft cloth.  

 

CARE INSTRUCTIONS

Proximity

®

 sensing technology works by sensing changes in the electrical field.

Nearby conductive materials that are electrically charged or not properly electrically grounded may interfere with the desired function of the

product.  If the controller is unable to calibrate, this is most likely due to poor isolation of the sensor, or improper grounding of nearby

conductive material.
To ensure best performance, it is important that these nearby electrical effects are minimized as much as possible.
For hardwire installations, it is essential for the controller to be grounded to an electrical ground.  Please note that supply or drainage 

plumbing may not be grounded to an electrical ground.  It is not always necessary to ground battery product, but in some installations this

may prove beneficial.
Cosmetically chromed drain plumbing may have non-conductive protective coatings, and may have electrical interruptions and should not

be used for grounding purposes.
Conductive sinks; often stainless or cast iron are a common source of electrical interference.  Such sinks should be grounded to an electrical

ground.  There may be protective coatings that insulate the sink, proper electrical connection should be confirmed.  On stainless sinks, this

may be achieved by connecting to the sink mounting hardware with the grounding strap.
Furthermore, it is essential that the sensor module and faucet are electrically isolated from the sink.  Be sure to use all supplied spacers.

There are spacers supplied for both the above and below deck surfaces.  The supplied spacers should aid in preventing the faucet shank

from contacting the sink, however care must be taken to ensure that the faucet, shank and sensor module are isolated from all nearby

conductive material.
Other items such as conductive counters, sink supports, or nearby power/data lines may also create interference and should be avoided/

routed away from or where possible, electrically grounded.

 

GROUNDING AND ISOLATION

Summary of Contents for 62 TP3 Series

Page 1: ...xx8TR Description Battery Operated Surface Mount Box 0 Battery Operated Control Box 1 Trim Battery Operated Control Box 8 Trim Hard Wire Operated Surface Mount Box 0 Hard Wire Operated Control Box 1 Trim Hard Wire Operated Control Box 8 Trim Page 2 4 6 8 10 12 COVER 060073A Screws 4 pkg and driver bit for covers 060577A 12 Stainless Steel Cover TMV Repair Kits 061137A Adjustment Wrench 063164A Che...

Page 2: ... in accordance with local plumbing and electrical codes FLUSH ALL PIPES THOROUGHLY BEFORE INSTALLATION Fig 3 w w w s p e c s e l e c t c o m 621TPxxx0 Battery Operated Surface Mount Box 0 Mount faucet to sink using the provided components in the following order grey bottom spacer washer lock nut Fig 1 NOTE The black top spacer must be used at all times between the spout and countertop sink and the...

Page 3: ...tance the ability to store an electrical charge is brought on by a 2nd conductor a human body entering the Proximity field As the human hand enters and leaves the vicinity of the electrical field around the spout the solenoid is opened and closed accordingly It is critical in the installation process that all electrical and mechanical connections be as tight and rigid as possible ensuring a smooth...

Page 4: ... at the box outlet Connect water supply through to spout Assure supply lines are completely flushed and free of debris Install aerator Product supplied as shown by solid lines All items shown by dotted lines supplied by others Typical Installation Recessed Mount Box Flexible Sensor Cord Conduit Polymer Braided Hose 257mm 10 13 Control Box 305mm 12 Stainless Steel Cover Plate 102mm 4 355mm 14 max F...

Page 5: ...round a conductor the spout and monitoring the change in capacitance that the conductor undergoes The change in capaci tance the ability to store an electrical charge is brought on by a 2nd conductor a human body entering the Proximity field As the human hand enters and leaves the vicinity of the electrical field around the spout the solenoid is opened and closed accordingly It is critical in the ...

Page 6: ... spacer is for temporary use only for flushing of system Must be replaced with solenoid and washers Figure 13 STEP 4 FLUSH SYSTEM SET TEMPERATURE Remove coverplate from control box Open screwdriver stop s to flush installation for one minute minimum 4A Run water for a sufficient time so the hot and cold water supplies are as hot and cold as possible 4B Place a thermometer in a plastic container an...

Page 7: ...tion is complete and the faucet is ready to use The cover should be secured at this time STEP 6 TEST FOR OPERATION Test for operation If OK then close lid Use caution not to damage wires or components on electronic driver board Route wires through notches openings in the box See Fig 18 Secure lid using screws If faucet leaks from spout outlet SHUT OFF WATER SUPPLIES Check proper solenoid connectio...

Page 8: ...e converter Refer to Fig 24 and Fig 25 for reference wiring diagrams Install the grounding kit assembly 061259A per the supplied instruction sheet for proper grounding with Conductive and Non Conductive sinks Turn on power supply for the transformer 5 quick beeps 3 times will be generated when power is first applied to the unit At this stage step back at least 3 feet from the controller and spout ...

Page 9: ... change in capacitance the ability to store an electrical charge is brought on by a 2nd conductor a human body entering the Proximity field As the human hand enters and leaves the vicinity of the electrical field around the spout the solenoid is opened and closed accordingly It is critical in the installation process that all electrical and mechanical connections be as tight and rigid as possible ...

Page 10: ...f system Must be replaced with solenoid and washers Figure 28 STEP 3 CONNECT WATER SUPPLY Install sink and connect drainage to rough in See Figure 28 Polymer braided hose supplied must be connected to the spout Other connection tubes and fittings are supplied by the installer to connect to the 1 2 nominal sweat at the box outlet Connect water supply through to spout Assure supply lines are complet...

Page 11: ...s The solenoid will stay open while hands are moving then close once hands are removed from the sensing zone CONDUCTIVE SINKS NON CONDUCTIVE SINKS Fig 33 Fig 32 Metal Sink Mounting Hardware or Bracket Control Box 1 Driver Board P3 GND 120VAC Ground Electrical Box Metal Box Ground Wiring Sink Ground Wiring Earth Ground Wiring Control Box 8 To Additional Sinks Driver Boards Vitreous China Sink Contr...

Page 12: ...cer is for temporary use only for flushing of system Must be replaced with solenoid and washers Figure 35 STEP 4 FLUSH SYSTEM SET TEMPERATURE Remove coverplate from control box Open screwdriver stop s to flush installation for one minute minimum 4A Run water for a sufficient time so the hot and cold water supplies are as hot and cold as possible 4B Place a thermometer in a plastic container and ho...

Page 13: ...LIBRATION After power is applied to the driver board it generates 5 quick beeps 3 times which prompts the installer to step away from the unit During the calibration process the water will turn ON for several seconds then proceed to shut off An additional 5 beeps indicates calibration is complete and the faucet is ready to use The cover should be secured at this time STEP 6 TEST FOR OPERATION Test...

Page 14: ...to store an electrical charge is brought on by a 2nd conductor a human body entering the Proximity field As the human hand enters and leaves the vicinity of the electrical field around the spout the solenoid is opened and closed accordingly It is critical in the installation process that all electrical and mechanical connections be as tight and rigid as possible ensuring a smooth signal is carried...

Page 15: ...seconds 2 0 2 5 3 0 minutes Metering Mode b Buzzer For Proximity the buzzer setting does not silence the controller and sensor status annunciations Buzzer options are ON and OFF the default setting is ON ON ON OFF M This feature changes the product into a metering faucet it will run the full set amount of time after the user s hands are first detected Note Auto Timer T is disabled when the feature...

Page 16: ...nd water closets HARSH ABRASIVES or POLISH To clean simply wipe the surface with a damp cloth and blot dry with a soft cloth CARE INSTRUCTIONS Proximity sensing technology works by sensing changes in the electrical field Nearby conductive materials that are electrically charged or not properly electrically grounded may interfere with the desired function of the product If the controller is unable ...

Page 17: ...oid are functioning Ensure inlet screen on solenoid is not plugged with debris Check the incoming water supply Turned on proper pressure and volume etc Check for a pressure lock between the mixing valve and the solenoid Turn off the water supply and undo the connection to the inlet of the solenoid this will eliminate any built up pressure Then activate sensor and test fire the solenoid If after ab...

Page 18: ...is warranty gives you specific legal rights and you may also have other rights which vary from state to state This warranty is for commercial products only from Delta Faucet Company and Delta Faucet Canada a division of Masco Canada Limited and is void for any damage to this faucet due to misuse abuse neglect accident improper installation any use in violation of instructions furnished by Delta Fa...

Reviews: