background image

 

 

Maintenance 

Keep the glass dome clean using clean water, with mild detergent or Isopropyl Alcohol for difficult deposits. 
If any two of the three blue desiccant indicators turn pink - unscrew the desiccant capsule from the indicator cap, 
and replace with a fresh one (type SPN1-SD). Recalibration every 2 years is recommended. 

SPN1-MS1 Specifications  

The following accuracy figures give 95% confidence limits, i.e. 95% of individual readings will be within the stated 
limits under normal climatic conditions. 
 

 

Spectral response of the Sunshine Pyranometer 
(thermopile, diffusers and dome combined),  
shown with the solar spectrum at ground level. 

SPN 1 Sp ectral resp onse

0

20

40

60

80

100

120

0

500

1000

1500

20 00

2500

3000

W avelength nm

rel

at

ive co

n

tr

ib

u

ti

o

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

Solar
SP N1

Overall accuracy: Total 
(Global) radiation and 
Diffuse radiation 

±

5% Daily integrals 

±

5% 

±

10 W.m

-2 

  Hourly averages 

±

8% 

±

10 W.m

-2 

  individual readings 

Resolution 0.6 

W.m

-2 

 = 0.6mV 

Range 

0 to >2000 W.m

-2

Analogue output sensitivity 

1mV = 1 W.m

-2

Analogue output range 

0-2500mV 

Sunshine status threshold 

200 W.m

-2

 in the direct beam  

Other specifications 

Accuracy: Sunshine status 

±

10% sun hours  

with respect to the threshold 

Accuracy: Cosine Correction 

±

2% of incoming radiation 

 over 0-90° Zenith angle    

Accuracy: Azimuth angle  

±

 5% over 360

°

 rotation     

 

Temperature coefficient 

±

 0.02 % /

°

C typical  

Temperature range 

-20 to + 70

°

C  

Stability 

Recalibrate every 2 years. 

 

Typical cosine response of the Sunshine Pyranometer 
compared to the ideal cosine curve. 
The upper curve shows the relative accuracy. 

S P N 1   C o sin e   re sp o n s e

0

0 .1

0 .2

0 .3

0 .4

0 .5

0 .6

0 .7

0 .8

0 .9

1

1 .1

1 .2

0

1 0

20

3 0

40

50

6 0

70

8 0

90

Ze nith   a ng le   (deg ree s)

Ide al c osine
A bs olu te  outpu t
Relative  error
2 %   limits

Response time 

< 200ms 

Spectral Response 

400-2700nm  

Spectral sensitivity 

10% typical 

Non-linearity <1% 
Tilt response 

negligible 

Zero Offsets 

< 3W.m

-2

 for a change of 5

°

C/hour 

in ambient temperature. 
<3 W.m

-2

 dark reading

 

Latitude capability 

-90° to + 90

°

 

Environmental: Sealing 

IP67  

Sunshine status output 

No sun = open circuit 
Sun  = short circuit to ground 

Power requirement 

2mA (excluding heater power),   
5V – 15V DC 

Heater power 

12V – 15V DC, up to 1.5 A  

Heater control 

Continuously variable up to 20W 
output for external temperatures 
below 0°C 

Lowest snow & ice-free 
temperatures  
(using internal heater) 

-20

°

C at  0 m/s wind speed 

-10

°

C at 2 m/s wind speed       

Mounting options: 

3 x M5 tapped holes in base;  
108mm pcd, 120°spacing 

Size & Weight 

140mm dia x 100mm high, 786 gm 

Notices 

Patents

: The SPN1 Sunshine Pyranometer is protected by patents EP1012633 & US 6417500. 

EMC certification

: available on request. 

 

 

Delta-T Devices Ltd

  

 

 

128 Low Road, Burwell, Cambridge,CB25 0EJ, UK

 

 

Tel:  +44 1638 742922  Fax: +44 1638 743155 

email: 

[email protected]

  web: 

www.delta-t.co.uk

 

All manuals and user guides at all-guides.com

Reviews: