Technical reference
Specifications
Technical Specifications for PR2/4 and PR2/6
Measurement
Volumetric soil moisture content,
θ
V
(m
3
.m
-3
or %vol.).
Range
Accuracy specified from 0 to 0.4 m
3
.m
-3
,
Full range is from 0.0 to 1.0 m
3
.m
-3
after soil specific
calibration
±
0.04 m
3
.m
-3
, 0 to 40°C
Accuracy
with generalised soil
calibration in 'normal' soils
±
0.06 m
3
.m
-3
, 0 to 40°C
Salinity errors
Included in above figures (50 to 400 mS.m
-1
)
Vertically: ~95% sensitivity within ±50mm of upper ring
of each pair.
Soil sampling
volume
Horizontally: ~95% sensitivity within a cylinder of radius
100mm.
0 to 40°C for full accuracy specification,
Environment
–20 to 60°C full operating range.
IP67 rated
Stabilisation
Full accuracy achieved within 1s from power-up.
Minimum: 5.5V DC with 2m cable, 7.0V with 100m.*
Maximum: 15V DC
Power
requirement
PR2/4 consumption: < 80 mA
PR2/6 consumption: < 120 mA
4 (PR2/4) or 6 (PR2/6) analogue voltage outputs:
Outputs
~0 to 1.0V DC corresponding to 0 - 0.6 m
3
-3
m
(mineral
calibration)
Cable
8-core screened. Maximum length: 100m
Construction
material
25.4mm polycarbonate tube with pairs of stainless steel
rings
PR2/4 length: 750mm
Weight: 0.6kg
Size / weight
PR2/6 length: 1350mm
Weight: 0.95kg
*using cables supplied by
Δ
T
26
z
Technical reference
Profile Probe User Manual
3.0a