Portable monitoring
Set up the HH2 meter
Connect the
Profile Probe
to your HH2
using the supplied
PRC/d-HH2
cable
Press
Esc
to turn the meter on, and if
necessary press again until it displays:
Delta-T Devices
Δ
TMoisture Meter
Make sure the meter is set to read from a PR2:
Press
Set
and scroll down to the
Device
option.
o:
.
Press
Set
again and scroll down t
Press
Set
to confirm this choice
Device:
v
PR
2
If you intend to store readings, you may find it useful to define
each reading by setting a plot label and sample number –
accessed by pressing
Set
and scrolling to the
Data
option.
See Calibration section for advice on setting
Soil Type
and
Soil Set-Up.
For other options, refer to the HH2 User Manual.
Taking readings
Insert the
Profile Probe
into an access tube.
Press
Read
to take a reading - it takes about 3 seconds.
Press the arrow keys to view
readings from other depths. You
can choose different units from
the
Display
option.
PR
2
Store
?
22.7%vol
v
100mm
Press
Store
to save or
Esc
to discard the reading.
If you want to maximise the sampling volume, take 2 further
readings with the probe rotated through 120° each time.
Remove the PR2, replace the access tube cap and move on to a
new site….
Viewing stored readings
If you have saved data, connect the HH2 to your
PC and run HH2Read to retrieve the readings.
12
z
Operation
Profile Probe User Manual
3.0a