- 60 -
9. TECHNICAL SPECIFICATIONS
Sensor features
Relative humidity RH (for HD45
17…, HD46
17… and HD46
17B… models)
Sensor Capacitive
Measuring range
0...100 % RH
-40…+85°C Dewpoint Td
Sensor operating range
-40...+80°C
Accuracy
±
2% (10..90%RH) @ 20°C,
±
2.5% in remaining range
See table for Dewpoint.
Resolution 0,1%
Temperature influence
2% over the whole temperature range
Hysteresis and repeatability
1% RH
Response time (T
90
) <20
sec.
(air speed = 2m/sec and constant temperature)
Long-term stability
1%/year
Temperature T (for HD45
17…, HD45
7B…, HD46
17… and HD46
17B… models)
Sensor type
NTC 10K
Ω
Measuring range
-30...+85°C (-22…+185°F)
Accuracy
(except models with current output)
±
0.2°C
±
0.15% of reading within 0…70°C
±
0.3°C
±
0.15% of reading within -30…0°C, 70…85°C
Accuracy
(for models with current output)
±
0.5°C
±
0.15% of reading within -30…+85°C
Resolution 0.1°C
Response time (T
90
)
<30 sec. (air speed = 2m/sec)
Long-term stability
0.1°C/year
Carbon Dioxide CO
2
(for HD45
7B…, HD45
B… and HD46
17B… models)
Sensor
NDIR double wavelength
Range
0…5000 ppm
Sensor operating range
-5…50°C
Accuracy
±
(50ppm+3% of reading) @ 20°C and 1013hPa
Resolution 1ppm
Temperature influence
0.1%f.s./°C
Response time (T
90
) <120
sec.
(air speed = 2m/sec and constant temperature )
Long-term stability
5% of measure/5 years
Summary of Contents for HD45 17A
Page 10: ... 10 3 1 HD45 connection diagrams ...
Page 11: ... 11 3 2 HD46 connection diagrams ...
Page 62: ... 62 Dimensions in mm HD45 series HD46 series ...
Page 63: ... 63 Fixing holes ...