13
If no temperature probe or combined probe with temperature sensor is connected, the in-
strument displays a dotted line (- - - -).
13)
YEAR
: to set the current year. Use the arrows to modify this parameter and confirm using
ENTER.
14)
MNTH (month)
: to set the current month. Use the arrows to modify this parameter and con-
firm using ENTER.
15)
DAY
: to set the current day. Use the arrows to modify this parameter and confirm using
ENTER.
16)
HOUR
: to set the current hour. Use the arrows to modify this parameter and confirm using
ENTER.
17)
MIN
: to set the current minutes. In order to correctly synchronize the minute, it is possible
to reset the seconds by pressing the °C/°F key. Use the arrows to set the current minute plus
one, and as soon as that minute is reached press °C/°F: this synchronizes the time to the
second. Press ENTER to go onto the next item.
18)
BAUD_RATE:
indicates the frequency used for the serial communication with the PC.
Values from 1200 to 38400 baud. Use the arrows to modify this parameter and confirm us-
ing ENTER.
The communication between instrument and PC (or serial port printer)
only works if the instrument and PC baud rates are the same
. If the USB connection is
used this parameter value is automatically set (please see the details on page 33).
LOG/DUMP LOG key - only HD2106.2
In measurement mode, this function starts and stops the logging of a data block to be saved in the
instrument's internal memory. The data logging frequency is set in the
"Print and log interval"
menu parameter. The data logged between a start and subsequent stop represent a block.
When the logging function is on, the LOG indication is displayed, the battery symbol blinks and a
beep is issued each time a logging occurs;
the battery symbol does not appear when using an
external power supply
.
To end the logging, press LOG.
If the Auto-HOLD function is enabled (please see the menu), the data logging is disabled.
The HD2106.2 can turn off during logging between one capture and the next: the function is con-
trolled by the
Sleep_Mode_LOG
parameter. When the logging interval is less than one minute, the
logging instrument remains on; with an interval of at least one minute, it turns off between one cap-
ture and the next if the parameter
Sleep_Mode_LOG
=YES.
>>>
DumpLOG - only HD2106.2
When the LOG key is pressed after the MENU key, the transfer of the logged data on the serial port
is started.
Please see the paragraph dedicated to data transfer on page 33.