background image

-      - 

15

C

ONDUCTIVITY MEASUREMENT

 

The instruments work with conductivity/temperature combined probes, 4-ring and 2-ring, only con-
ductivity probes, or temperature probes. The 4-wire Pt100, 2-wire Pt1000 probes may be used for 
measuring temperature, which is used for the automatic compensation of the conductivity. 
The instrument obtains the following from the measurement of conductivity:  

 

the liquid resistivity measurement (

Ω

, k

Ω

, M

Ω

),  

 

the concentration of total dissolved solids (TDS) according to the 

χ

/TDS conversion factor, 

which can be modified using the menu, 

 

the salinity (NaCl quantity in the solution, expressed in g/l).  

The conductivity, resistivity, TDS or salinity indication is displayed in the main line, while the sec-
ondary line shows the temperature. 
The conductivity probes must be periodically calibrated. To facilitate this operation, four automatic 
calibration solutions recognized by the instrument are provided:  

 

0.001 Molar KCl solution (147

μ

S/cm @25°C),  

 

0.01 Molar KCl solution (1413

μ

S/cm @25°C),  

 

0.1 Molar KCl solution (12880

μ

S/cm @25°C),  

 

1 Molar KCl solution (111800

μ

S/cm @25°C),   

 

User calibration of the temperature sensors is not required. 

 

The probes are detected during turn on, and this cannot be performed when the instrument is 
already on, therefore if a probe is connected and the instrument is on, it must be turned off 
and back on again

Standard probe 

The standard 4-ring conductivity/temperature combined probe has code 

SP06T

A bell made of Pocan delimits the cell measurement zone. A positioning key, present in the probe's 
end part, orients the bell correctly when the probe is introduced. For cleaning, simply pull the bell 
along the probe's axis without rotating it. 

It is not possible to perform measurements without 

this bell. 

 

 

This probe's temperature measuring range is 0…+90°C. 
 

4- ring or 2- ring probes  

The HD2106.1 and HD2106.2 conductivity meters use 4-ring or 2-ring probes for conductivity 
measurement.  
The 4-ring probes are preferred to measure high conductivity solutions, either over an extended 
range or in presence of pollutants. The 2-ring probes operate in a shorter measurement range but 
with accuracy comparable with the 4-ring probes.  

Summary of Contents for HD2106.1

Page 1: ...written in this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manual cann...

Page 2: ...26 INSTRUMENT SIGNALS AND FAULTS 27 LOW BATTERY WARNING AND BATTERY REPLACEMENT 29 INSTRUMENT STORAGE 30 SERIAL INTERFACE AND USB 31 STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER 33 THE LOGGING...

Page 3: ...ransferred from the instrument connected to a PC via the RS232C serial port or USB 2 0 port The storing interval printing and baud rate can be configured using the menu The HD2106 1 and HD2106 2 model...

Page 4: ...4 Conductivity meter Thermometer HD2106 1...

Page 5: ...etween the current value and the logged value when the key is pressed in the menu decreases the current value 9 SERIAL key starts and ends the data transfer to the serial communication port 10 MENU ke...

Page 6: ...6 Conductivity meter Thermometer HD2106 2...

Page 7: ...the serial communication port In the menu clears the data contained in the instrument s memory 10 LOG DUMP LOG key during normal operation starts and ends the saving of the data in the internal memor...

Page 8: ...ally it sets the instrument ready for normal measurement MIN cm min s 2 3 m m k During turning on should no probes be connected the last manually set temperature appears in the secondary line The unit...

Page 9: ...n of this MENU key below Attention the data captured using the Record function cannot be transferred to the PC CAL key It increases the current parameter when used in the menu when used in measurement...

Page 10: ...ment of memorized data only HD2106 2 the message _LOG_DUMP_or_ERAS Transfer data or erase is scrolled in the comment line The center figure reports the number of free memory pages FREE Pressing SERIAL...

Page 11: ...ero Pressing SERIAL only causes the print to occur when the measurement has stabilized HOLD symbol still Later it is possible to re peat the print at will but while the HOLD mode is on the sample iden...

Page 12: ...conversion factor depends on the nature of the salts present in the solution in the field of water quality treatment and control where the main component is CaCO3 Calcium Carbonate a value of 0 5 is u...

Page 13: ...and PC baud rates are the same If the USB connection is used this parameter value is automatically set please see the details on page 33 LOG DUMP LOG key only HD2106 2 In measurement mode this functio...

Page 14: ...when using an external power supply the battery symbol does not appear Press SERIAL to end the continuous printing Before starting the printing with SERIAL set the baud rate To do so select the Baud R...

Page 15: ...1413 S cm 25 C 0 1 Molar KCl solution 12880 S cm 25 C 1 Molar KCl solution 111800 S cm 25 C User calibration of the temperature sensors is not required The probes are detected during turn on and this...

Page 16: ...low constant K probes are preferably used for low conductivity values the high constant ones for high values The indicative measurement range is reported in the following diagram S cm mS cm 0 1 0 1 0...

Page 17: ...ation deletes the data of the previous calibration For example if the instrument is calibrated in 4 points and a new 1 point calibration is performed the instrument will be calibrated in 1 point only...

Page 18: ...end the probe calibration press TDS ESC 9 Rinse the probe with water If you are then going to perform low conductivity measurements we recommend rinsing the probe using distilled or bidistilled water...

Page 19: ...nt as it was before the calibration 13 Rinse the probe with water If you are then going to perform low conductivity measurements we recommend rinsing the probe using distilled or bidistilled water The...

Page 20: ...0 124 1173 10 72 94 4 27 0 153 1467 13 37 115 7 17 0 126 1199 10 95 96 3 28 0 157 1494 13 62 117 7 18 0 128 1225 11 19 98 2 29 0 161 1521 13 87 119 8 19 0 130 1251 11 43 100 1 30 0 164 1548 14 12 121...

Page 21: ...ensor is housed in the end part of the probe When measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip probe In order to per...

Page 22: ...ctivity cell are welded directly onto the connector s pins On the left there are the 1 4 points on which the Pt100 or Pt1000 sensor wires must be welded The JP1 JP4 jumpers are in the center of the ca...

Page 23: ...ns 1 and 4 Jumper between pins 2 and 5 4 electrode conductivity probe Internal view of the connector None As an alternative the Pt100 and Pt1000 sensors can be welded directly on the connector s pins...

Page 24: ...nnector Jumper between pins 1 and 4 Jumper between pins 2 and 5 4 ring conductivity probe Pt100 temperature sensor Pt100 Internal view of the connector None Ensure the welds are clean and perfect Once...

Page 25: ...be held perpendicular against the surface Apply oil or heat conductive paste between the surface and the probe in or der to improve contact and reduce reading time Whatever you do do not use water or...

Page 26: ...Check that the cell is not dirty that there are no air bubbles inside it Calibrate again using the appropriate standard Slow response or instability Check that the cell is not dirty that there are no...

Page 27: ...ory full the instrument cannot store further data the memory space is full CAL blinking No calibration has yet been effected or the cell constant value has been modified in the menu please see the des...

Page 28: ...OG logging in progress LOG MEM FULL memory full LOG_CLRD memory data cleared LOG_DUMP_OR_ERAS transfer or erase data LOG_STOP logging complete LOG_STRT logging started MIN USE_UNIT_TO_ZERO SEC minutes...

Page 29: ...lace the batteries in order to turn the instrument back on In order to avoid data loss the logging session is ended if the HD2106 2 is logging and battery voltage falls below the minimum operating lev...

Page 30: ...to direct sunlight The instrument may be exposed to a source of high temperature The instrument may be exposed to strong vibrations The instrument may be exposed to steam salt or any corrosive gas Ma...

Page 31: ...m Cable Instrument port PC port Installation of USB drivers HD2110CSNM RS232 MiniDin RS232 9 pole SubD No C 206 RS232 MiniDin USB Yes CP23 USB Mini USB USB No The instrument standard serial transmissi...

Page 32: ...the temperature n 0 C n 1 F LUBn Sets the unit of measurement for the conductivity n 0 S cm n 1 n 2 TDS n 3 NaCl RA Reading of LOG PRINT interval set RP 600 Battery level Resolut 0 01V RUA U C Temper...

Page 33: ...upon pressing the LOG key the instrument logs the first data and turns off 15 seconds before the next logging instant it turns on again to capture the new sample and then turns off The data stored in...

Page 34: ...a standard serial communication program e g Hyperterminal only through the RS232 serial port of the instrument using the cable HD2110CSNM or the cable C 206 Connect the HD40 1 printer using cable HD21...

Page 35: ...rument must be switched off 2 Using the Delta Ohm HD2110CSNM or C 206 cable connect the measurement instrument to the first free RS232C COM or USB serial port of the PC 3 Turn on the instrument and se...

Page 36: ...safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product...

Page 37: ...2Vdc 1A Security of memorized data Unlimited independent of battery charge conditions Time Date and time Schedule in real time Accuracy 1min month max drift Measured values storing model HD2106 2 Type...

Page 38: ...0 999 cm 1 cm 1 00k 19 99k cm 0 01k cm 20 0k 99 9k cm 0 1k cm 100k 999k cm 1k cm 1 10M cm 1M cm Measurement range Kcell 10 Resolution 0 5 5 0 cm 0 1 cm Accuracy resistivity 0 5 1digit Measurement of t...

Page 39: ...conversion factor 0 4 0 8 Cell constant K cm 1 0 01 0 1 0 7 1 0 and 10 0 Standard solutions automatically detected 25 C 147 S cm 1413 S cm 12880 S cm 111800 S cm TECHNICAL DATA OF PROBES WITH MODULE I...

Page 40: ...C 4 electrode cell Pocan Platinum SPT401 001 K 0 01 0 04 S cm 20 S cm 0 120 C 2 electrode cell AISI 316 steel SPT01G K 0 1 0 1 S cm 500 S cm 0 80 C 2 electrode cell Glass Platinum SPT1G K 1 10 S cm 1...

Page 41: ...h in built temperature sensor RCT The kit includes 4 thermal paper rolls 57mm wide and 32mm in diameter HD22 2 Laboratory electrode holder composed of basis plate with incorporated magnetic stirrer st...

Page 42: ...tor length 2 metres TP47 1000 O Immersion probe sensor Pt1000 Probe s stem 3mm length 230mm 2 wire connection cable with connector length 2 metres TP87 100 O Immersion probe sensor Pt100 direct 4 wire...

Page 43: ...ali e nazionali garantita da una catena di riferibilit che ha origine dalla taratura dei campioni di prima linea dei laboratori accreditati di Delta OHM presso l Istituto Primario Nazionale di Ricerca...

Page 44: ...2002 96 EC the European users of electrical and elec tronic equipment can return it to the dealer or manufacturer upon purchase of a new one The il legal disposal of electrical and electronic equipmen...

Reviews: