5
Display Description
When you turn the instrument on the display appears as in the figure.
The display is backlit: To enable or disable the backlight, select the menu item Menu >> System
Parameters >> Backlight.
You can adjust the contrast level: Use the
arrows to modify it.
The display is laid out as described below:
1.
Battery
symbol
: It indicates the battery level.
2.
EPT
symbol
: The
EPT
(End PoinT) symbol indicates the display mode. The mode selection is
done using the
EPT
key. Until the EPT is blinking, the measurement is updated on the
display; when it is fixed, the measurement is “frozen”. For a new measurement, press MEAS.
EPT =
DIR
: The instrument operates in
continuous view mode
. In this mode the displayed
measurement is updated every second (standard mode).
EPT =
MAN
: The displayed measurement is continuously updated until MEAS is pressed.
During the measurement update, the MAN symbol is blinking. For a new measurement, press
MEAS.
EPT =
TIME
: The measurement is frozen after a set time of 8 seconds. For a new
measurement, press MEAS.
EPT =
AUTO
: The instrument performs the measurement, and when it stabilizes the AUTO
symbol stops blinking. For a new measurement, press MEAS.
In the following figure you can see an example of the measurement process with the
EPT
AUTO
function enabled. After setting the EPT = AUTO function using the EPT key, the
electrode is immersed into a liquid. To perform the measurement, press MEAS. the AUTO