Delta OHM HD 2020 Manual Download Page 9

-      - 

The  HD 2020 calibrator allows calibrating any sound level meter  provided that it is equipped with a labora-
tory or wo

rking standard 1/2” microphone compliant with  IEC 61094-01 and IEC 61094-4 standards. 

Corrections depending on the microphone 

The    HD2020  calibrator  generates  a  94  dB  (or  114  dB)  sound  pressure  level  referred  to  20

μPa.  Working 

standard 

½”  microphones  for  sound  level  meters  are  manufactured    to  achieve  flat  frequency  response  in 

free or diffuse field, i.e. in a field of progressive plane waves propagating in the same direction as the micro-
phone axis and in a field of sound waves coming from every direction, respectively. These propagation con-
ditions are different from those in the calibrator cavity.  
In  free  field,  reflections  due  to  the  microphone    alter  the  sound  level  by  increasing  the  high-frequency  car-
tridge  effective  sensitivity.  Microphones  optimized  for  free  field  measurements  exploit    this  phenomenon  to 
achieve  flat  frequency  response  even  at  very  high  frequencies.  In  these  microphones,  the  sound  level  in-
crease at 1 kHz corresponds to approximately 0.05 dB ÷ 0.20 dB. Therefore, when you calibrate a free field 
microphone, you must allow for this  difference when you set a sound level 0.1 dB or 0.2 dB lower than the 
calibrator nominal one.  
Microphones optimized for diffuse field measurements 

don’t require corrections when calibrated  in a closed 

cavity at 1 kHz instead.

 

 
 
 
 

 

 

L

OW BATTERY INDICATION AND REPLACEMENT

 

 
The HD 2020 calibrator is provided with two batteries: a 9V alkaline battery that can be replaced by  the user 
and a lithium buffer battery. The latter allows the date indicator and watch to work even without external bat-
tery: it must be replaced by a Delta Ohm authorized service centre.  

The 9V battery power is constantly monitored:  

  If the battery is fully charged, its symbol is off;  

  If the battery is partially charged, its symbol blinks: please replace the battery as soon as possible; 

  If the battery is uncharged and the instrument cannot work properly, its symbol is constantly on. When the 

battery is uncharged, the calibrator switches off  in about 10 seconds. 

 

To replace 9V batteries, switch off the instrument and open the lid at the bottom of the instrument. Replace 
the battery. Close the lid.  

Date, time and calibration interval are stored if the buffer battery is charged. 
The average life of the buffer battery depends on the external battery: if you use it, the average life of the in-
ternal battery is  about 5 years.  
 
  

Battery tips 

 

Remove the battery if the instrument is not to be used for an extended period. If the battery  is low, 

replace it as soon as possible. 

 

Make sure that there is no loss of liquid. 

 

Use good quality sealed batteries (alkaline if possible). 

 
 
 
 
 

 

Summary of Contents for HD 2020

Page 1: ...on written in this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize Data figures and descriptions contained in this manual cannot be legally asserted We reserve the right to make changes and corrections without prior notice REV 1 1 15th June 2012 HD 2020 ENGLISH ...

Page 2: ...he standard screen 3 Display When you turn the instrument on it shows all segments on then the sound pressure level 94 or 114dB will appear in standard view 4 Calibrator cavity for 1 2 inch microphones 5 key in standard mode it selects 94dB and 114dB pressure levels alternately In menu mode it de creases the current value 6 key in standard mode it selects 94dB and 114dB pressure levels alternately...

Page 3: ...st Le misure effettuate presso un Laboratorio di Taratura Accredia sono garantite da una catena di riferibilità ininterrotta che ha origine dalla taratura dei campioni di prima linea del Laboratorio presso l istituto metrologico nazionale Measurements performed in an Accredia Calibration Laboratory are guaranteed by a uninter rupted reference chain which source is the calibration of the Laboratory...

Page 4: ... is inserted the instrument switches off 5 minutes after turning on provided that you don t press any key The calibrator display shows calibration pressure level battery life current date and time The internal clock allows to set up the validity period of calibration expressed in years and months starting from the date of adjustment as soon as the time interval has elapsed a blinking symbol alerts...

Page 5: ... after about 5 minutes if the microphone is inserted in the cavity and you don t press any key If the cavity is open the sound level blinks on the display in this case the instrument switches off automati cally after 30 seconds You cannot disable the AutoPowerOff function SETUP Key The SETUP key allows entering and viewing the menu To exit press the SETUP key until you go back to the standard scre...

Page 6: ...rent time Use the arrow keys to increase decrease the value Press the SETUP key to confirm and go to the current minute Use the arrow keys to increase decrease the value Press the SETUP key to confirm and go to the current second Whenever you press the arrow keys seconds will be reduced to zero Press the SETUP key to confirm and go to calibration section of the menu the first triangle on the left ...

Page 7: ...s entered by the calibration laboratory and the expiry date When the time interval expires the triangle blinks to remind you that the calibrator needs a new ad justment years of validity second triangle Use the arrow keys to increase decrease the value Confirm with SETUP key and go on to the number of va lidity months months of validity second triangle Use the arrow keys to increase decrease the v...

Page 8: ... and the calibrator off switch on the sound level meter and detect the unweighted ambient sound pressure level If the measured level is below 78 dB you can use both calibration sound levels 94 dB and 114 dB if the level is between 78 dB and 98 dB you can use the 114 dB level only while a sound level above 98 dB means that calibration is not possible Press the ON OFF key to switch on the instrument...

Page 9: ...ibrator nominal one Microphones optimized for diffuse field measurements don t require corrections when calibrated in a closed cavity at 1 kHz instead LOW BATTERY INDICATION AND REPLACEMENT The HD 2020 calibrator is provided with two batteries a 9V alkaline battery that can be replaced by the user and a lithium buffer battery The latter allows the date indicator and watch to work even without exte...

Page 10: ...the battery compartment The display and the keypad take up the higher part The electro acoustic transduction device is on the left and it consists of a wide cavity with a piezoceramic generator and a feedback sensor The device emits the signal through the microphone An outer capillary hole balances the chamber static pressure protecting microphones from overpressure due to their insertion Control ...

Page 11: ...erated by the piezoelectric transducer through automatic gain control The signal provided by the sensor slightly varies with the ambient temperature and the static pressure the frequency of the signal provided by the oscillator is also stable in relation to ambient parameters INSTRUMENT STORAGE Instrument storage conditions Temperature 25 55 C Humidity below 90 RH without condensation When storing...

Page 12: ...al functioning and operating safety are guaranteed only if climatic conditions are the same as described in the specifications chapter Use and store the instrument avoiding Sudden change of the ambient temperature that may cause condensation Inflammable or corrosive gases Direct vibration or blows Intense electromagnetic fields static electricity If you carry the instrument from a cold to a hot en...

Page 13: ... 10 C 50 C and 10 90 RH Static pressure influence 0 1 dB in the range 65 kPa 108 kPa Stability levels Short term stability 0 03 dB Stability after 1 year normal use 0 1 dB Operating conditions Working temperature 10 50 C Relative humidity 90 R H Storage temperature 25 70 C Microphone equivalent volume 5 to 250 mm 3 Power supply 9V alkaline battery IEC type 6LR61 9V rechargeable batteries are also ...

Page 14: ... 14 ORDERING CODE HD2020 The kit includes HD2020 calibrator one 9V alkaline battery instruction manual and I SO 9001 calibration report ...

Page 15: ...N 4 DISPLAY AND KEYPAD DESCRIPTION 5 CALIBRATION PROCEDURE 8 LOW BATTERY INDICATION AND REPLACEMENT 9 CONSTRUCTION AND FUNCTIONING 10 INSTRUMENT STORAGE 11 FUNCTIONING AND OPERATING SAFETY 12 SPECIFICATIONS 13 ORDERING CODE 14 ...

Page 16: ...he European users of electric and electronic devices are allowed to give back to the Distributor or Manufacturer the used device at the time of purchasing a new one The illegal disposing of electric and electronic devices is punished by a pecuniary administrative penalty This guarantee must be sent together with the instrument to our service centre N B Guarantee is valid only if coupon has been co...

Reviews: