ENGLISH
96
KEY
SIMBOLS LIT BESIDES
THE NUMBERS
FUNCTION
DESCRIPTION
P11+
=
calibrate tempe-
rature probe
pressing the PROG key moves on
to step P11.
- P11 -
Pressing the PROG key moves
on to step P12.
When ENTER is pressed with
P11 on the display, the instrument
enters the program for calibrating
the temperature probe.
If a temperature probe (4-wire
Pt100 probe or series TP 870) is
inserted in input B, the instrument
will calibrate this probe, initially
setting the default calibration
values.
If there is no temperature probe
inserted in input B, the instrument
will calibrate the temperature pro-
be present in the dissolved
Oxygen measurement probe
inserted in input A, initially setting
the default values.
In the program for calibrating the
temperature probe, the instrument
shows the message CAL at the
top of the display and, at the bot-
tom of the display, the temperatu-
re value of the probe currently
being calibrated.
For greater details, see the instru-
ment temperature calibrating
instructions.
°
C