![DELTA M microtuf FS3200 Series Operation And Maintenance Manual Download Page 30](http://html1.mh-extra.com/html/delta-m/microtuf-fs3200-series/microtuf-fs3200-series_operation-and-maintenance-manual_2478262030.webp)
DELTA M Corporation
3242-OM-04 3/27/02
26
7.0 SPECIFICATIONS
TYPE:
Thermal Differential-Dual RTD Sensors
PROCESS CONNECTIONS:
0.75” (19.05mm) MNPT Standard, 0.5"(12.7mm), 1" (25.4mm)
MNPT, and various flanges optional.
INSERTION LENGTH:
Two inch (50.8mm) Standard, (shorter 0.5 inch (12.7mm) and
longer to 120 inch (3048mm) optional).
CONSTRUCTION MATERIALS:
Wetted parts are 316L SS welded construction (alternate
materials for corrosive environments available as options.
Consult factory.)
AGENCY INSTRUMENT RATINGS:
CSA Explosion Proof: (CS series) T4A
Class I, Group B, C, and D
Class II, Group E, F, and G
CENELEC/Explosion Proof (CN Series):
EEx d IIC T4(Akron Electric Enclosure)
EEx d IIB T4(Killark Enclosure)
CE: EMC Directive: 89/336/EEC (CE Option)
OPERATING TEMPERATURE:
Process: -70°C to + 200°C (-100°F to +390°F) standard
(to + 600°C (+1000°F) optional
Electronics: -40°C to +60°C (-40°F to +140°F)
PRESSURE RATED:
To 3000 psig (20.4 MPa)
RANGE
Gaseous Mass Flow: 0.1 to 500 fps
Liquid Mass Flow: 0.01 to 5 fps
REPEATABILITY:
± 1% of Set Point or ± 1/32 inch (±.8mm)
TIME RESPONSE:
0.5 to 10 seconds no-flow (dry) to flow (wet) and 2 to 60
seconds flow (wet) to no-flow (dry) (application dependent)
INPUT POWER:
115 Vac, 50/60HZ standard. (230 Vac, 50/60HZ, 24 Vdc, or 24
Vac optional); 3.1w. maximum.
DELTA M PART NO.
FUSE REQUIREMENTS (F1):
CSA/FM CENELEC
115 Vac: 1/4 amp
101603
101605
230 Vac: 1/4 amp
101603
101605
24 Vdc: 1/4 amp
101603
101605
OUTPUT:
5A, 250 VAC, DPDT Standard (Optional 10A, 250 Vac SPDT)
5A 30 VDC
STABILITY:
Temperature compensated over entire range.