Expert
Logger
4.4.2.3
XML network configuration template
The XML template allows you to set the network configuration for a device,
without the device having to be connected to a PC. This is particularly help-
ful if you have multiple devices to configure, as you can then modify the
file for each device and save it to a USB memory stick. When the USB
memory stick is connected to a device, this file is read and the device
obtains the settings from the file.
Method
1. Save the file to your PC (right click on the link).
2. Open the file using an editor.
3. Change entries
DefaultGateway=’192.168.0.254’, IPV4_
Address=’192.168.0.1’
and
IPV4_NetMask=’255.255.255.0’
to your
required values or change
DHCP_active=’false’
to
DHCP_active=’true’
.
4. Save the amended file to a USB memory stick.
5. Plug the USB memory stick in to the device.
6. Switch on the device.
4.4.2.4
USB network driver
When connecting the device to a computer via the USB interface, the required
driver is automatically installed.
Connecting to the device
Enter the IP address of the device in DataService Configurator. If you are
not sure about the address:
1. Open the Network and Sharing Center of your PC via Control Panel.
2. Click
Change adapter settings
.
39
Delphin Technology AG
Januar 2018
Expert
Logger
Manual