is used with the DMC program and the ElastiMeter. See the user manual of the 

DMC for further instructions.  

 

Device info 

Displays the software version, the manufacturing date, the serial number of the 

device, the serial number of the programmed receiver unit and the measurement 

counter. 

 
 
R

ECOMMENDATIONS FOR MEASURING WITH THE 

E

LASTI

M

ETER

 

 

Skin surface 

 

o

 

Smooth and even skin site is the most reliable measurement site. 

o

 

Skin hair, scars and folds may prevent a good contact between the probe 

and the skin.  

o

 

Skin surface should be dry.  

o

 

Avoid skin sites over superficial bones, tendons and large veins. 

 

Environmental conditions 

 

The  ElastiMeter  can  be  used  in  normal  room  temperature.  Recommended 

operating conditions are mentioned in the section 

Technical data

 in this manual. 

 

Test person related factors 

Measurements  with  the  ElastiMeter  can  be  performed  when  the  subject  has 

been resting for 5-10 minutes. Hot drinks and smoking should be avoided for 

at least 60 min before measurements. 

o

 

Make  sure  that  the  tested  person  is  totally  relaxed  before  the 

measurement. No muscle tension should occur.  

o

 

The tested person should be in the same position at every measurement 

session  during  a  study  (e.g.  sitting  on  the chair  the  arm  resting  on  the 

table, if an arm is measured). 

 

 

 

 

 

Summary of Contents for ElastiMeter

Page 1: ...www delfintech com ElastiMeter User Manual...

Page 2: ...FACE DESCRIPTION 4 OPERATING INSTRUCTIONS 5 ADJUSTING SETTINGS 7 RECOMMENDATIONS FOR MEASURING WITH THE ELASTIMETER 8 MAINTENANCE CLEANING AND DISINFECTING 9 CALIBRATION 10 TECHNICAL DATA 11 APPENDIX...

Page 3: ...2...

Page 4: ...and thus the skin s response under a short term load indicates its instant elastic properties APPLICATIONS Instant skin elasticity measurements give important information on the biophysical propertie...

Page 5: ...indicator 3 Battery voltage indicator 4 Display 5 Number of the sample 6 Instant skin elasticity ISE value 7 Past values ElastiMeter Accessories available ElastiMeter measuring unit DMC data collectio...

Page 6: ...5 seconds o The indenter touches the skin first then the base plate o The indenter imposes a constant deformation when the base plate is in full contact with the skin o The force sensors recognize the...

Page 7: ...ISE reading is the mean of 5 consecutive measurements Readings under the recent value are means of previous measurements If the probe is left on the skin for too long the message OFF SKIN is displayed...

Page 8: ...stiMeter on by pressing the operating button briefly 2 When the device is on press the button again and hold it until the MENU is displayed Then release 3 The indicator starts to scroll downwards thro...

Page 9: ...o Avoid skin sites over superficial bones tendons and large veins Environmental conditions The ElastiMeter can be used in normal room temperature Recommended operating conditions are mentioned in the...

Page 10: ...d with ethanol Set the moistened soft paper on the table and wipe the measurement probe slightly to the paper Use only soft paper tissue with no lint The probe head must be completely dry before use N...

Page 11: ...Do not use rechargeable batteries 4 Replace and screw the bottom cover in place Do not touch the circuit board with any object CALIBRATION Delfin ElastiMeter is recommended to be calibrated every oth...

Page 12: ...x 45 x 35 mm o Weight with batteries 153 g o Protection class IP52 o Measurement completion time ISE 0 5 seconds o Measurement repeat cycle ISE 10 measurements per minute typically o Measured paramet...

Page 13: ...Storage at temperature 5 C to 35 C o Storage at relative humidity 5 to 80 Warranty o Limited two year warranty see Appendix C Bluetooth module certifications o Contains FCC ID QOQBLE113 o Contains IC...

Page 14: ...s not correct or the probe is left on the skin for too long LO FORCE Applied contact force too low HI FORCE Applied contact force too high OK FORCE Applied contact force correct LO SPEED Skin contact...

Page 15: ......

Page 16: ...contact Make sure the measuring surface base plate is fully and evenly in contact with the skin Make sure your hand is supported during the measurement and the base plate touches the skin perpendicula...

Page 17: ......

Page 18: ...not apply to any defect failure or damage caused by improper use or inadequate maintenance and care Delfin Technologies shall not be obligated to furnish service under this guarantee 1 to repair dama...

Page 19: ...Delfin Technologies Ltd info delfintech com www delfintech com...

Reviews: