
PRO-WEI100-MAN-19V01 (0225)
Copyright © 2019 Define Instruments
16
values to alter the current calibration.
CALSET allows the user to select and switch between calibration sets, giving them the option
of saving and restoring a previous calibration.
Auto calibration
B _ _ _ CAL SET X – CALIBRATE ZERO
scrolls across the display, where
X
is the Cal
Set where the current calibration will be stored (see 6.3M).
– CALIBRATE ZERO
toggles with the currently selected option. Use the
and
buttons to select
YES
or
NO
, and then press
.
¨
If you selected
YES
, continue to 6.3C now.
¨
If you selected
NO
, skip to 6.3D now.
C _ _ _ REMOVE WEIGHT – PRESS P TO ACCEPT
scrolls across the display and
toggles with the current no-load value. Remove the weight. Use the
and
buttons to adjust the no-load value if required, and then press
.
D _ _ _ CALIBRATE SPAN
scrolls across the display and toggles with the current
selection. Use the
and
buttons to select
YES
or
NO
, and then press
.
¨
If you selected
YES
, continue to 6.3E now.
¨
If you selected
NO
, skip to 6.3F now.
E
_ _ _ ADD CAL WEIGHT – ENTER DESIRED SPAN – PRESS P TO ACCEPT
scrolls
across the display and toggles with the current span value. Apply a calibration
weight to the weigh platform. Then use the
and
buttons to adjust the
value, and press
to accept.
F
If Auto calibration was successful, you will be directed out of the calibration
menu to the operational display without viewing any further scrolling messag-
es. (To enter step 6.4, you must select
SKIP
at 6.3A.)
If calibration fails,
_ _ _ CALIBRATION FAILED
will scroll across the display and
you will be directed back to the operational display. Check your signal and con-
nections, and then repeat the calibration procedure.