3. Attach the Frame to the Backboard.
Without interrupting manual CPR, position the Frame over the patient.
Attach the Frame to the Backboard by aligning the Frame latches over the Backboard
pins and pushing down until the latches snap into place. The latches may be clicked into
place one at a time or simultaneously.
Pull up on the Frame to make sure it is securely attached to the Backboard.
WARNING
If patient is too large for the Frame, remove Frame and continue manual CPR compressions.
Do not use the Lifeline ARM if the Frame cannot be latched to the backboard.