background image

5

DAC-U2519EN-BB

Setup Instructions

4.1  Assembly

WARNING

Use only the Wall Charger supplied with the Charging 

Stand. Using other chargers or cables may cause unsafe 

conditions.

1.  Place the Charging Stand near an accessible AC power outlet on a level 

surface.

2.  Ensure that the Wall Charger plug adapter (if supplied) is correct for your 

AC power outlet.

3.  Connect the Wall Charger cable to the Charging Stand by inserting the 

Conncetor Jack into the Charger Port until it clicks into place and locks 

(see 

Figure 1

 at left). 

4.  Plug the Wall Charger into an AC outlet. The LED on the Wall Charger 

and the Charging Stand will momentarily light green.

 

4.2  Charging the Battery Pack 

It takes about 3 hours to charge a new fully-depleted Battery Pack. The charging 

time differs depending on operating environment.

1.  Before inserting any Battery Pack into the Charging Stand, ensure that 

the date is within the Battery Pack’s useful life specifications.

2.  Insert the Battery Pack into the Battery Pack Slot, pushing firmly until 

the Battery Pack clicks into place.

3.  Upon initial insertion, the charging process begins immediately as 

indicated by a blinking green LED on the Charging Stand and the Wall 

Charger. Throughout the charging process, these LEDs will continue to 

blink green.  

Note: 

 If the LED indicator does not light, check all cables 

and connections. 

4.  When charging is complete, the LED on the Charging Stand and the 

Wall Charger will be steady green.

Note: 

 Refer to the Troubleshooting section of this document should other indicators 

be present or if all connections are good but the LED indicators do not illuminate.

4.3  Removing the Battery Pack After Charging 

For ease of maintenance, it is recommended that the Battery Pack should be fully 

charged before putting it back into use unless it is required for emergency use.   

1.  Press the release latch on the Charging Stand to release the Battery 

Pack from the Charging Stand. 

2.  Confirm that the Battery Pack is sufficiently charged by inserting it into 

the AED.  See the appropriate AED User Manual for additional details. 

3.  If the Battery Pack does not have sufficient capacity, ensure that 

another battery is available or see Section 5,

“Troubleshooting.”

Note: 

Leaving a Battery Pack in the Charging Stand after it has reached a fully-

charged state will not damage the Battery Pack or the Charging Stand.

Summary of Contents for DBP-2009

Page 1: ...DAC U2519EN BB Defibtech DBP 2009 Automated External Defibrillator AED Rechargeable Battery Pack and Charging Stand User Manual ELECTRONIC DISTRIBUTION...

Page 2: ...sed in the examples are fictitious unless otherwise noted Limited Warranty The Defibtech Limited Warranty is the sole and exclusive warranty for Defibtech defibrillators and associated accessories Ref...

Page 3: ...y Pack After Charging 5 4 4 Post Charging and Storage 6 4 5 After Each Use 6 5 Troubleshooting 6 5 1 LED Indications 6 5 2 Troubleshooting Table 7 6 Disposal 8 6 1 Notice to European Union Customers 8...

Page 4: ...DAC U2519EN BB...

Page 5: ...ed to deal with the constraints of periodically removing the rechargeable battery pack from the AED and charging it as compared to others that prefer the longer life battery of a non rechargeable batt...

Page 6: ...ables may cause unsafe conditions WARNING Do not charge a Battery Pack which is damaged discolored or has leakage It may cause overheating explosion or fire WARNING Select a location where the Wall Ch...

Page 7: ...Do not drop the Battery Pack or Charging Stand or apply a mechanical shock to it CAUTION Do not store or use the Battery Pack or Charging Stand outside of the specified conditions CAUTION Before clea...

Page 8: ...LE BATTERY PACK LED Indicator CHARGING STAND Battery Pack Slot Charger Port Battery Release Latch WALL CHARGER LED Indicator Connector Release Connector Figure 1 Component Description DBC 2001 Chargin...

Page 9: ...tion the charging process begins immediately as indicated by a blinking green LED on the Charging Stand and the Wall Charger Throughout the charging process these LEDs will continue to blink green Not...

Page 10: ...After Each Use Follow the Defibtech AED User Manual instructions on the recommended steps after each use or as otherwise indicated including performing a battery insertion test which assesses battery...

Page 11: ...Clean the Charging Stand contacts Replace the Battery Pack with a new one Replace the Wall Charger with a new one The Battery Pack temperature is out of range Put the Charging Stand including Wall Cha...

Page 12: ...this product can only be disposed of in compliance with the provisions of the above mentioned European directive and following possible revisions as well as with the corresponding national regulation...

Page 13: ...any connectors 8 Maintenance The Charging Stand and Wall Charger are designed for minimal maintenance When the Charging Stand and Wall Charger have not been recently used dropped or otherwise as requi...

Page 14: ...lled in AED Visual Audible Environment Operating Conditions Temperature 0 to 50 C 32 to 122 F Humidity 5 to 95 non condensing Environment Storage and Transport Conditions Temperature 0 to 50 C 32 to 1...

Page 15: ...32 to 104 F Humidity 15 to 95 non condensing Environment Storage and Transport Conditions Temperature 20 to 80 C 4 to 176 F Humidity 15 to 95 non condensing Safety Standard IEC 60601 1 2012 Type of p...

Page 16: ...storage requirements See environmental requirements Keep dry X X X X Do not damage or crush Do not expose to high heat or open flame Do not incinerate Y Y Y Y Operational temperature limitation This...

Page 17: ...per disposal procedures Manufacturer Serial number Lot number yy ww Y Y Y Y Catalogue number 11 1VDC 2 0 Ah 22 2 Wh Li Ion 3ICR16 65 Lithium ion battery Y Y Y Y Use by yyyy mm IPX0 No special protecti...

Page 18: ...ston Post Road Suite 201 Guilford CT 06437 USA Tel 1 866 333 4241 Toll free within North America 1 203 453 4507 Fax 1 203 453 6657 Email sales defibtech com Sales reporting defibtech com Medical Devic...

Reviews: