Loading a Sample

 

 

1.

 

Sample Preparation:

 

Þ

 

Liquid Sample: Pull 5mL of the desired liquid into 5mL syringe.

 

Þ

 

Soil Sample: Place 1g soil in sparge bottle. (An application note for Methanol Extraction can be found on our website.)

 

2.

 

Place the load/analyze valve in the LOAD position.

 

3.  Slip bottle over green tube (sparge needle). 

 

4.

 

Attach syringe to the instrument and transfer the sample to the sparge bottle.

 

Þ

 

For Soil Sample: Fill the syringe with 5mL deionized water, then attach it and transfer the water to sparge bottle containing 

the soil sample. 

 

5.

 

Attach sparge bottle in the UP position and gently tighten the sparge bottle nut counter clockwise. 

 

6.

 

Place load/analyze valve in the ANALYZE position. 

 

  

Running a Sample

 

  

1.

 

Instrument screen reads “READY”

 

2.

 

Record Log # and other relevant sample data for future reference. 

 

3.

 

Press the start button in Ellvin or on the instrument to begin an analysis cycle.

 

4.

 

As the analysis cycle proceeds, the instrument screen displays the analysis steps.

 

5. If connected to Ellvin, a chromatogram of results is displayed on the computer screen. 

 

6.

 

If not connected to Ellvin, the instrument screen displays a chromatogram when the cycle is finished.

 

7.

 

If not connected to Ellvin, press the instrument start button to scroll through analyte concentration estimates. 

 

8.

 

Before loading/running another sample,

 

a.

 

Empty the Sparge bottle.

 

b.

 

Wait for the instrument screen to read “READY”.

 

NOTE: Samples must be run from low to high concentrations.  

 

Rinse the instrument and create a clean baseline between runs if concentrations are unknown.

 

 

VOC ANALYZER 

 

GC TEMP: 40C  

READY

 

LOG_016

 

BTEX V. 965 2012

 

1, 2

 

3

 

or

 

 

CLEANING 4

 

 

BENZE     33PPB   

48.33S

 

7

 

4

 

5

 

 

6

 

1

 

2

 

3a

 

4

 

LOAD 

 

ANALYZE

 

5

 

3b

 

Summary of Contents for FROG-4000

Page 1: ...wer 4 Turn the instrument power switch to the ON position fully up The instrument display screen will be visible 4 Open Ellvin by double clicking on the Ellvin icon 5 Ellvin opens in the default Live...

Page 2: ...position Running a Sample 1 Instrument screen reads READY 2 Record Log and other relevant sample data for future reference 3 Press the start button in Ellvin or on the instrument to begin an analysis...

Page 3: ...files Open the folder and the screen displays a list of sequential files named LOG_XXX 3 Select the LOG file of interest and drag copy it to the computer 4 Open Ellvin 5 Open Live Data Window 6 Click...

Page 4: ...Look In to select the FOLDER labeled with the DATE that Ellvin created converted the file of interest 4 Click on the file of interest and a preview chromatogram appears on the right side of the scree...

Page 5: ...t is CONNECTED and whether or not parameters have been updated GC Column Temperature Indicator Displays the GC temperature The icon is GREEN when instrument is ready for loading The icon is RED when t...

Page 6: ...of the features in Live Data Window the following features are available in Analyze Window Integrator 1 The user creates a baseline and integrates peaks then calculates the peak height and area to det...

Page 7: ...he analyte in the Analyte Alias window maximum of 5 Characters 3 Click the ADD button Correlation Chart and Equations This feature in the Calibration Window displays the degree of correlation between...

Page 8: ...s Tb Ramp time from GC cold to GC hot temperature seconds Tc Hold time at hot GC temperature seconds Ct Initial cold GC temperature o C Ht Final hot GC temperature o C CLEAN Time cleaning PC by heatin...

Page 9: ...air sampler by connecting to this port Figure 5 Connect exhaust connector to exhaust port ANALYZE Position Figure 3 Sample Introduction Valve Coupler Line Transfer Line Sparge Needle Port Valve Contr...

Page 10: ...ampler can now be connected to a Tedlar bag to calibrate the instrument or to the wand to collect air samples 5 Press the yellow start button on the instrument or in Ellvin to begin an air sample anal...

Page 11: ...position is facing toward you 5 Turn valve position counterclockwise to the right place sparge bottle in up position on Frog 4000 and tighten the nut around sparge bottle 6 Run Frog 4000 as normal by...

Page 12: ...d high drain rate battery packs Neither Defiant Technologies nor Tenergy Corporation are responsible for damage caused by charging other battery pack brands with this charger Use 1 8A current for batt...

Page 13: ...Sparge Bottle 3 Frog With Sparge Tube Protector 4 Serial Data Cable 5 USB Cable to N 6 Battery 7 Air Sampler a Transfer Line b Air Sample Couple Line c 1 16 3 16 Adapter 8 Power Cable 9 Battery Charg...

Page 14: ...rge Bottle 3 Frog With Sparge Tube Protector 4 Serial Data Cable 5 USB Cable to N 6 Battery 7 Air Sampler a Transfer Line b Air Sample Couple Line c 1 16 3 16 Adapter d Air Sample Wand 8 Power Cable 9...

Reviews: