Cleaning water tank
To keep the water tank clean, the water used should be pure water,
distilled water or cold boiled water. we suggest to change water every day.
The water tank should be cleaned every week. First use mild detergent
and shake the tank to wash. Then use clean water to rinse it,
After cleaned the water tank, please remember to put water tank plastic
plug back to avoid leakage of oxygen from the water inlet, caused no
oxygen output from oxygen outlet mouth when the machine is in normal
working.
Air filter cleaning and replacement
The white air filter at the back of the machine can reduce noise, isolate
dust and purify the air. It should be cleaned around every ten days. First
use mild detergent to wash, and then rinse it with clean water. Pay special
attention that it must be dry to be positioned back, or the function of the
molecular sieve could be damaged, which will lead to decreased oxygen
concentration.
Cleaning oxygen Inhaler and atomizing components
Oxygen inhaler should be cleaned after use, and the nasal suction head
on the nasal inhaler should be cleaned and disinfected after use.
Soaking the inhaler in the solution of 5% potassium ferric acid for 5
minutes, cleaning with water, or wiped with alcohol. Oxygen tube as
consumable should be replaced regularly.
Fuse Replacement
After confirmed the power cord plug is out of power suppler outlet,using a
slotted screwdriver to pry off the fuse holder cap, take out the defective
glass fuse tube ,replace new one, put the cap back and tighten it.
NO.
1
Phenomenon
Cause analysis
Solution
Odor in air.
2
Water flows from the
oxygen outlet.
3
4
5
The machine lelse
"puff" sound every
several seconds
After startup, the
machine operates
well, the flowmeter
displays "0" or is
too small
After startup, there
is bubble in the
water tank,butthere
is no oxygen
getting out from the
oxygen outlet.
In the initial stage of using,
there is smell for the
connection tube and other
plastic parts. It is normal.
The water level of the
humidifier water tank is beyond
the highest water level, so the
airflow pressed the bubble into
the oxygen tube.
The machine has been
used for a long time, so that
the steam in the airflow
gathers on the tube wall.
lt is the sound of nitrogen
discharging which is normal
and is asign of fine working.
The flowmeter is off or the
value istoo small
The oxygen tube is twisted or
blocked.
The inner connection tube of
the machine is disconnected or
broken.
The silica gel stopper of the
humidifying water tank is not
properly positioned.
Turn up the oxygen flow.
Straighten the oxygen tube.
Disconnect the powercord and
contact professionals to repair it.
Remove the humidifying water
tank to plug the silica gel
stopper.
The humidifier water tank
hasn't been cleaned for a long
time or the water hasn't been
changed.
Clean the oxygen tube and
humidifier water tank.
Keep the water lower than the
highest water level.
pour the water in water tank.
And add water into the water
tank when there is no water
coming out from oxyhen outlet.
The odor that comes from
silicone tube and ABS rubber
parts is normal and non-toxic by
inspection.The odor will
disappear as time goes by.