3.2.3. Detector Control Unit
Table 3.5:
Detector Control Unit Ratings
Power input
1 x 100 VAC to 240 VAC, 50
/
60 Hz, 1 A to 3 A, 250 W (Platinum)
Dimensions (W x H x D)
434
.
0 mm x 42
.
8 mm x 495
.
0 mm
Weight
<
10
.
6 kg
Chassis
1U
3.2.4. Thermal Stabilization Unit
Table 3.6:
Thermal Stabilization Unit Ratings
Power input
Single phase 100 VAC to 240 VAC, allowable voltage range
±
10 %, 50/60 Hz, 8 A (100 VAC) to 3 A (240 VAC)
Dimensions (W x H x D)
270 mm x 393 mm x 436 mm
Weight
17
.
5 kg
Typical flow
3 L min
−
1
Maximum operation pressure
3 bar
3.3. Ambient Conditions
The PILATUS3 R CdTe 100K-M detector is equipped with a temperature and a humidity sensor. When either sensor detects
that the operating conditions are not met, the detector will shut off. However, as the sensors may not prevent damage,
temperature and humidity should be monitored to avoid breaching the operation limits.
Information
#2
The relative humidity within the module chamber must be lower than 30 % during operation and lower than
25 % during start up (use of dry air or nitrogen advised).
The PILATUS3 R CdTe 100K-M detector is designed for indoor use only. The ambient conditions shown in table 3.7 must
be satisfied. The stated values are for the ambient conditions.
Values inside the detector, in particular due to the dry-air or nitrogen supply, are different. These are described in section 5.4
and chapter 6.
Table 3.7:
Detector operating ambient conditions
Ambient Condition
Value
Operating temperature
+20
◦
C to +35
◦
C
Operating humidity
<80 % at + 20
◦
C, non-condensing
Storage temperature
+15
◦
C to +40
◦
C
Storage humidity
<40 % at +20
◦
C, non-condensing
Please consider following points when storing the detector:
PILATUS3 R CdTe 100K-M Technical Specifications v1.0.0
7 | 29