DECKMA HAMBURG GmbH
Issue: 27.05.08
Instruction Manual OMD-2005
Page 14 of 25
10.3 Functional
Tests
a) Run oil free water through the instrument to purge the system.
b) Adjust the flow rate through the unit by using the small screws in the cell cap
(Fig. 1, Pos. 2). Taking out a screw will increase the flow rate.
NB
: The flow rate should be checked on both, the clean water supply and the
separator sample supply. If the clean water supply is obtained from a high
pressure source, the flow rate will be higher than from the sample point.
The flow rate is not influencing the accuracy of the instrument. The adjustment
is only important for the time delay between the sample point and the monitor.
c) Switch on the instrument and make sure, that the Power LED is illuminated
and the display is
showing the initializing
display for about 15
sec. After that time it
will change to the
standard display,
showing the actual measurement.
d) During oil free water is running through the monitor check the Zero adjustment
according Section 11. The display should be "0" to “2” and the status will show
“FW”. If the display varies by greater amounts, it may be that air entrainment is
present. If this is the case, the cause must be located and rectified.
f) If the Zero need to be adjusted, this can be done in the programming mode as
described in section 10.4. (Service – Offset)