INSTALLATION
4.1 Location
The Chloralert™ should ideally be mounted outside the area to be monitored. However it can be mounted in the
contaminated area if it is considered absolutely necessary. The location should be such that the ambient temperature
never falls below -20 °C (-4°F) nor rises above 65 °C (149°F).
It should never be mounted in direct sunlight and should be protected from the weather if mounted outside.
4.2 Mounting
Securely mount the Chloralert™ at a convenient working height (approx. 1,3 - 1.5 m from the floor).
4.3 Sampling line
As the Chlorine weight is higher then the air, we recommend that the sample inlet point is located at
about 200 mm (8") from the floor. Use of frigid or flexible 1” PVC tubing is recommended. It is
recommended to install the proper insect screen (supplied) at the terminal point of the sample line to
avoid dust suction.
WARNING
NEVER USE TUBES WITH A DIAMETER LOWER THAN 1" WHEN
THE SAMPLE PIPE LENGTH EXCEEDS 4 M
Figure 3 - Typical Installation
CHLORALERT
R
E
S
E
T
A
L
A
R
M
T
S
T
E
WHEN
FLASHING
C
C
N
R
E
V
E
S