Pressure Module, M-P and Pressure Temp Module, M-PT
Document No. 800 1013-1
27
Calibration:
1.
Connect the pressure transducer with the pressure manometer to the P3 connector. Select
Calibrate P3. Leave the transducer to room air pressure.
2.
Press the ComWheel to start zeroing.
3.
Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
4.
Set the pressure on the display to match the pressure reading on the manometer and press
the ComWheel.