![Datex-Ohmeda S/5 M-P Technical Reference Manual Slot Download Page 27](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-p/s-5-m-p_technical-reference-manual-slot_513846027.webp)
Pressure Module, M-P and Pressure Temp Module, M-PT
Document No. 800 1013-1
23
4.2
Troubleshooting flowchart
Figure 8
Troubleshooting flowchart
Possibly faulty
M-P/PT module
Take all modules out
of the Frame
Insert M-P/PT and turn
the power on
Does fault
still appear?
Fault not in M-P/PT
module
No
Module-ID
on screen?
Does another
module work
in the same slot?
Yes
No
No
Check front panel key
functions in Service
menu
Do they work?
Yes
Check front panel unit
and PT board to
find culprit. Replace if
necessary
Yes
No
Has EEPROM
failed
Has Temp
measurement
failed?
Has InvBP
measurement
failed
No
No
Yes
Yes
Yes
Recalibrate the temp
channels
Yes
Enter the Service menu
Replace the PT board
Replace the PT board
Replace the PT board
Recalibrate the
InvBP channel
OK ?
OK ?
No
No