S/5 Anesthesia Record Keeping Keyboard and Barcode Reader
Document No. 8001019-3
15
5.1
Keyboard menu
A service menu for testing the command board functions and for setting the K-ARKB type.
NOTE: All counters are automatically zeroed on returning to the previous menu.
Upper Led
is for testing the upper alarm LED (red) on the
command board. When the text is highlighted, the upper
alarm LED can be turned on and off by pressing the
ComWheel.
Lower Led
is for testing the lower alarm LED (yellow) on
the command board. When the text is highlighted, the
lower alarm LED can be turned on and off by pressing the
ComWheel.
Dummy Press
is for testing the ComWheel. When the text
is highlighted, pressing of the ComWheel create a sound
from the loudspeaker and the corresponding number on
the service data field increase.
Service Data
Message count
counts the number of messages that are
sent out to the main CPU board.
Leds upper
and
lower
indicate the states of the alarm
LEDs on the command board.
Direct action keys
texts are indications to the command
board membrane keys. When a key on the command
board is pressed, the corresponding text in the menu
changes its colour.
Control wheel, Press
counts the ComWheel pressings.
Control wheel, Clockwise
and
Counterclockwise
the
ComWheel turnings.
Since
shows the date and the time of the last run time
reset.
Summary of Contents for S/5 B-ARK
Page 2: ......
Page 4: ...Datex Ohmeda S 5 monitors Document No 8001019 3 ii SERVICE CHECK FORM A 1...
Page 26: ...Datex Ohmeda S 5 monitors Document No 8001019 3 22...
Page 27: ...APPENDIX A Service check form K ARKB Document No 8001019 3 23 APPENDIX A...
Page 28: ...Datex Ohmeda S 5 monitors Document No 8001019 3 24...
Page 30: ...Datex Ohmeda S 5 monitors Document No 8001019 3...