Displays and Controller Boards
Document No. 8001004-1
25
3.
Check external parts:
−
the outer cover is intact
−
the display screen is intact
−
the front panel stickers are intact
−
the ComWheel cover is intact and is attached properly
−
the ON/STBY -switch and its protector are intact and are attached properly
−
the anesthesia keyboard connector is clean and intact
−
the LCD interface cable connector is clean and intact
−
the block screws for the cable are in place and are tightened properly (if installed)
−
the block screw threads are intact (if installed)
?
4.
Check that the ON/STBY -switch changes its state firmly when switching it back and forth.
Leave the switch into STBY -position.
?
•
Turn the monitor to stand-by. Install the M-NE(12)STPR/M-ESTPR/M-ESTP to the Central
Unit.
•
Connect and lock the LCD interface cable to the LCD display and to the used B-DISP/B-
DVGA in the monitor.
5.
Check that the stand-by LED on the LCD display front panel is lit up (the monitor power cord is
connected to the mains).
?
•
Turn the LCD display ON/STBY -switch to ON.
?
6.
Wait until normal monitoring screen appears onto the LCD display. Check that the picture on the
LCD display screen is clear and stable. Check also that all the colors are clear.
?
Summary of Contents for D-VHC17
Page 2: ......
Page 6: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor Document No 8001004 1 iv ...
Page 56: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 50 Document No 8001004 1 ...
Page 57: ...Appendices A G Displays and Controller Boards Document No 8001004 1 51 APPENDICES A G ...
Page 58: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 52 Document No 8001004 1 ...
Page 60: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor A 2 Document No 8001004 1 ...
Page 62: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor B 2 Document No 8001004 1 ...
Page 64: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor C 2 Document No 8001004 1 ...
Page 66: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor D 2 Document No 8001004 1 ...
Page 68: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor E 2 Document No 8001004 1 ...
Page 70: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor F 2 Document No 8001004 1 ...
Page 72: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor G 2 Document No 8001004 1 ...