11.
Rotation adjustment:
Press the key FUNCTION. Select the rotation adjustment by pushing the key SELECT (-/+).
Adjust the rotation by pushing the key ADJUST (-/+). Store the adjustment by pushing the
key STORE.
?
12.
Degaussing:
Press the key DEGAUSS. Check that the picture on the screen sways briefly with a subdued
sound
NOTE: Degaussing (demagnetisation) takes place automatically when the monitor is turned
on. The efficiency of a single degaussing phase decreases if repeated at intervals shorter
than 10 minutes.
?
13.
Color adjustments:
Press the key FUNCTION. Select the color adjustments by pushing the key SELECT (-/+).
Adjust the colors by pushing the key ADJUST (-/+). Store the adjustments by pushing the
key STORE.
?
14.
Perform electrical safety check and leakage current test.
?
15.
Check that the display functions normally after the performed electrical safety check.
?
16.
Clean the picture tube with a cloth dampened with isopropyl alcohol.
NOTE: Do not use water with any kind of detergent for cleaning the picture tube. Do not use
any abrasives or hard objects to remove stains.
Clean the display cover with a cloth dampened with a mild detergent solvent.
NOTE: Do not use strong solvents (e.g. acetone) or aerosol cleaners for cleaning the cover.
?
•
Fill in all necessary documents.
Summary of Contents for D-VHC17
Page 2: ......
Page 6: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor Document No 8001004 1 iv ...
Page 56: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 50 Document No 8001004 1 ...
Page 57: ...Appendices A G Displays and Controller Boards Document No 8001004 1 51 APPENDICES A G ...
Page 58: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor 52 Document No 8001004 1 ...
Page 60: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor A 2 Document No 8001004 1 ...
Page 62: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor B 2 Document No 8001004 1 ...
Page 64: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor C 2 Document No 8001004 1 ...
Page 66: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor D 2 Document No 8001004 1 ...
Page 68: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor E 2 Document No 8001004 1 ...
Page 70: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor F 2 Document No 8001004 1 ...
Page 72: ...S 5 Anesthesia Monitor and S 5 Critical Care Monitor G 2 Document No 8001004 1 ...